HPLC system LHLC-A11

This model of HPLC gradient system is equipped with 2 sets of high-pressure constant-current pump and ultraviolet detector. Its microprocessor reciprocating plunger pump comes with functions such as high working pressure, low pulse rate. This gradient HPLC model meet the needs of conventional analysis to micro flow analysis.

Request Quote
High pressure pump Two sets of pump
Flow rate range 0.001 - 9.999 mL / min (0.001 mL / min increments)
Pressure range ≤ 0.5 MPa (0 - 42 MPa)
Flow rate precision RSD < 0.06 %
Pulsation ≤ 0.1 MPa (flow rate 1 ml / min, Pressure 5 - 10 MPa)
Qualitative analysis RSD 6 ≤ 0.1 % (naphthalene methyl alcohol solvent)
Quantitative repeatability RSD 6 ≤ 0.3 % (naphthalene methyl alcohol solvent)
Flow rate setting error Ss ≤ 2 % (1 mL / min water, 5 - 10 MPa room temperature)
Air tightness of the pump Pressure 42 MPa, 10 min time, Pressure drop < 0.5 MPa
Mixing accuracy ± 1 %
Mixing precision ± 0.2 %
Size (W × H × D) 450 × 300 × 160 mm
UV Detector
Wavelength range190 - 680 nm
Flow cell8 µL
Spectrum band width8 nm
Baseline noise≤ ± 0.25 × 10-5 AU (empty cell)
Wavelength accuracy± 1 nm
Baseline drift≤ ± 0.4 × 10-4 AU (empty cell)
Wavelength precision< 0.1 nm
Limit of detection3 × 10-9 g / mL (naphthalene methyl alcohol solvent)
Size (W × H × D)450 × 300 × 160 mm
  • More precise optical system for the high precision and small offset
  • Improved light energy through the parallel double cone flow cell
  • Significant reduction in the static noise of the instrument and signal drift

This HPLC system has applications in the field of biochemistry, chemistry, pharmacology for chemical analysis, poison analysis and in protein food detection, medicine testing and polymer, environmental analysis.

Compare Models

Model High pressure pump Flow rate range Pressure range Flow rate precision Pulsation Qualitative analysis Quantitative repeatability Flow rate setting error Air tightness of the pump Mixing accuracy Mixing precision Size (W × H × D)
LHLC-A10 Single pump 0.001 - 9.999 mL / min (0.001 mL / min increment) ≤ 0.5 MPa (0 - 42 MPa) RSD < 0.06 % ≤ 0.1 MPa (flow rate 1 ml / min, Pressure 5 - 10 MPa) RSD 6 ≤ 0.1 % (naphthalene methyl alcohol solvent) RSD 6 ≤ 0.3 % (naphthalene methyl alcohol solvent) Ss ≤ 2 % (1 mL / min water, 5 - 10 MPa, room temperature) - ± 1 % ± 0.2 % 450 × 300 × 160 mm
LHLC-A11 Two sets of pump 0.001 - 9.999 mL / min (0.001 mL / min increments) ≤ 0.5 MPa (0 - 42 MPa) RSD < 0.06 % ≤ 0.1 MPa (flow rate 1 ml / min, Pressure 5 - 10 MPa) RSD 6 ≤ 0.1 % (naphthalene methyl alcohol solvent) RSD 6 ≤ 0.3 % (naphthalene methyl alcohol solvent) Ss ≤ 2 % (1 mL / min water, 5 - 10 MPa room temperature) Pressure 42 MPa, 10 min time, Pressure drop < 0.5 MPa ± 1 % ± 0.2 % 450 × 300 × 160 mm
LHLC-B11 High pressure constant-current infusion pump 0.001 - 9.999 mL / min (in 0.001 mL / min increments) - - - - - - - - - 260 × 130 × 420 mm
LHLC-B10 High pressure gradient pump 0.001 - 9.999 mL / min - ± 5 % - - - - - - - 520 × 420 × 160 mm