Temperature and Humidity Test Chamber LTHC –A11 is designed to simulate constant climatic conditions for testing a wide variety of products for their quality, performance, shelf life and stability. It's used to simulate constant temperature and humidity conditions, thereby creating a physiologically ideal environment for product testing. A good temperature humidity chamber is essential for evaluating potential success or failure of a product in the field.
Capacity | 80 L |
Temperature range | (- 20 °C) to 150 °C |
Low type | I – 40 °C, II – 70 °C |
Environmental condition | 5 ℃ ~ + 40 ℃ ≤ 85 % RH |
Humidity range | 20 % - 98 % RH |
Temperature fluctuation | ± 0.5 °C |
Temperature deviation | ± 2 °C |
Humidity deviation | ± 2.5 % RH |
Heating rate | 3 °C / min |
Cooling rate | 1 °C / min |
Temperature sensor | PTR Platinum resistance PT 100 Ω / MV A-class |
Temperature resolution | ± 0.001 ℃ |
Humidity sensor | Dry and wet bulb sensor |
Humidity resolution | ± 1 % RH |
Internal dimension | 400 × 400 × 500 mm |
Overall dimension | 900 × 1100 × 1000 mm |
Power supply | AC 220 , 50 Hz |
Extensively used in basic and advanced research, pharmaceutical industries, chemical industries, food and beverage industry.
Available Range :
Capacity : 50 L Capacity : 80 L Capacity : 100 L Capacity : 225 L Capacity : 500 L Capacity : 800 LRelated Models :
Model | Capacity | Temperature range | Low type | Environmental condition | Humidity range | Temperature fluctuation | Temperature deviation | Humidity deviation | Heating rate | Cooling rate | Temperature sensor | Temperature resolution | Humidity sensor | Humidity resolution | Internal dimension | Overall dimension | Power supply |
LTHC-A10 | 50 L | (- 20 °C) to 150 °C | I – 40 °C, II – 70 °C | 5 ℃ ~ + 40 ℃ ≤ 85 % RH | 20 % - 98 % RH | ± 0.5 °C | ± 2 °C | ± 2.5 % RH | 3 °C / min | 1 °C / min | PTR Platinum resistance PT 100 Ω / MV A-class | ± 0.001 ℃ | Dry and wet bulb sensor | ± 1 % RH | 320 × 350 × 450 mm | 800 × 1050 × 950 mm | AC 220 , 50 Hz |
LTHC-A11 | 80 L | (- 20 °C) to 150 °C | I – 40 °C, II – 70 °C | 5 ℃ ~ + 40 ℃ ≤ 85 % RH | 20 % - 98 % RH | ± 0.5 °C | ± 2 °C | ± 2.5 % RH | 3 °C / min | 1 °C / min | PTR Platinum resistance PT 100 Ω / MV A-class | ± 0.001 ℃ | Dry and wet bulb sensor | ± 1 % RH | 400 × 400 × 500 mm | 900 × 1100 × 1000 mm | AC 220 , 50 Hz |
LTHC-B10 | 100 L | (-20)℃ to 150 ℃ ,(-40) ℃ to 150 ℃ , (-70 )℃ to 150 ℃ | - | - | 20 % ~ 98 % R.H | - | ± 2.0 ℃ | ± 2.5 % RH | - | - | - | - | - | - | - | - | - |
LTHC-B11 | 225 L | (-20)℃ to 150 ℃ ,(-40) ℃ to 150 ℃ , (-70 )℃ to 150 ℃ | - | - | 20 % ~ 98 % R.H | - | ± 2.0 ℃ | ± 2.5 % RH | - | - | - | - | - | - | - | - | - |
LTHC-B12 | 500 L | (-20)℃ to 150 ℃ ,(-40) ℃ to 150 ℃ , (-70 )℃ to 150 ℃ | - | - | 20 % ~ 98 % R.H | - | ± 2.0 ℃ | ± 2.5 % RH | - | - | - | - | - | - | - | - | - |
LTHC-B13 | 800 L | (-20)℃ to 150 ℃ ,(-40) ℃ to 150 ℃ , (-70 )℃ to 150 ℃ | - | - | 20 % ~ 98 % R.H | - | ± 2.0 ℃ | ± 2.5 % RH | - | - | - | - | - | - | - | - | - |
LTHC-B14 | 1000 L | (-20)℃ to 150 ℃ ,(-40) ℃ to 150 ℃ , (-70 )℃ to 150 ℃ | - | - | 20 % ~ 98 % R.H | - | ± 2.0 ℃ | ± 2.5 % RH | - | - | - | - | - | - | - | - | - |
LTHC-B20 | 100 L | (-70 )℃ to 150 ℃ | - | - | 20 % ~ 98 % R.H | - | ± 2.0 ℃ | ± 2.5 % RH | - | - | - | - | - | - | - | - | - |
LTHC-B21 | 225 L | (-70 )℃ to 150 ℃ | - | - | 20 % ~ 98 % R.H | - | ± 2.0 ℃ | ± 2.5 % RH | - | - | - | - | - | - | - | - | - |
LTHC-B22 | 500 L | (-70 )℃ to 150 ℃ | - | - | 20 % ~ 98 % R.H | - | ± 2.0 ℃ | ± 2.5 % RH | - | - | - | - | - | - | - | - | - |
LTHC-B23 | 800 L | (-70 )℃ to 150 ℃ | - | - | 20 % ~ 98 % R.H | - | ± 2.0 ℃ | ± 2.5 % RH | - | - | - | - | - | - | - | - | - |
LTHC-B24 | 1000 L | (-70 )℃ to 150 ℃ | - | - | 20 % ~ 98 % R.H | - | ± 2.0 ℃ | ± 2.5 % RH | - | - | - | - | - | - | - | - | - |
LTHC-B30 | 3150 L | -40℃ to 150℃ | - | - | 10% to 98% RH | - | ± 2.0℃ | ± 2.5 % RH | - | - | PT-100 platinum resistance, A class, Temperature sensor | - | Dry and wet bulb sensor | - | - | - | 380 V, 50 Hz |
LTHC-C10 | - | - | - | - | - | - | - | - | - | - | - | - | - | - | - | - | - |
LTHC-C20 | 150 L | 10 to 70°C | - | - | 40 to 95% | - | - | - | - | - | - | - | - | - | - | - | 220 V/50 Hz |
LTHC-B25 | 1500 L | (-60) ℃ to +150 ℃ | - | - | 10% to 98% RH | - | ± 1.0 ℃ | ± 2.5 % RH | - | - | - | - | - | - | - | - | - |