| Generator Eluent Types | KOH/NaOH/LiOH/MSA |
| Eluent Concentration Range | 0.1 to 100mM |
| Eluent Concentration Increment | 0.1mM |
| Eluent Flow Rate | 0.1 to 5.0mL/min |
| Maximum Pressure | 30MPa |
| Minimum Pressure | 5MPa |
| Accuracy Of Gradient | 0.0015 |
| Vacuum Degasser Degree | -70kPa |
| Degasser Internal Volume | 300μL |
| Degasser Maximum Flow Rate | 10mL/min |
| Degassing Efficiency | 1.0mL/min 90% |
| Degassing Volume | 7.5mL |
| Alarm system | Pressure Alarm, Liquid Leakage Alarm, and Washing Liquid Alarm |
| Flow Path | Chemically inert, metal-free PEEK flow paths |
| Pump Type | High-pressure and low-pulse two-piston tandem advection pump |
| Pump Maximum Pressure | 42MPa |
| Pump Flow Rate | 0.001 to 9.999mL/min |
| Pump Pressure Ripple | ≤ 0.5% |
| Pump Flow Precision | ≤ 0.1% |
| Pump Flow Accuracy | ≤ 0.1% |
| External Diameter of The Tube | 1/16" |
| Valve Control Mode | Stepper Motor |
| Valve Power Supply | 24V (DC) |
| Column Heater Temperature Range | Ambient + 5 to 60 ℃ |
| Column Heater Temperature Accuracy | ±0.1℃ |
| Temperature Stability | ≤ 0.05℃/h |
| Columns Supported | 2, 3, 4, and 5 mm I.D. |
| Analytical Column Size | 250 mm |
| Guard Column Size | 50 mm |
| Conductivity Detector Type | Auto-range conductivity |
| Conductivity Detector Cell Volume | ≤ 0.8μL |
| Conductivity Detection Range | 0 to 50000μS/cm |
| Conductivity Detector Resolution | ≤ 0.0020nS/cm |
| Detector Output Voltage | -6000 to +6000mv(adjustable) |
| Detector Electronic Noise | 0.02nS |
| Detector Baseline Noise | ≤ 0.0002μS/cm |
| Detector Baseline Drift | ≤ 0.001μS |
| Detector Temperature Range | Ambient +5℃ to 60℃ |
| Detector Temperature Accuracy | ±0.01℃ |
| Detector Temperature Compensation | 1.7 %/℃ |
| Detector Maximum Pressure | 10.0MPa |
| Detector Instrument Linearity | ≥ 0.999 |
| Detector Quantitative Repeatability | ≤ 0.2% |
| Detector Qualitative Repeatability | ≤ 0.5% |
| Minimum Detectability Concentration | Clˉ ≤ 0.0001ug/mL Li⁺ ≤ 0.0001ug/mL |
| Suppressor Type | Self-Regeneration electrolytic micro-membrane suppressor |
| Suppressor Maximum Pressure | 6.0MPa |
| Non-Suppressed Chromatography | Yes/support |
| Suppressor Void Volume | <40μL (4mm) <15μL (2mm) |
| Suppressor Power Supply | 150W |
| External Dimension (W × L × H) | 360 × 500 × 560 mm |
| Net Weight | 26 Kg |
| Gross Weight | 32 Kg |
| Accessories | Quantity |
| Ion chromatograph host | 1 |
| High-pressure double plunger advection pump | 1 |
| Conductivity detector | 1 |
| Anion column | 1 |
| Anion guard column | 1 |
| Anion suppressor | 1 |
| Cation column | 1 |
| Cation guard column | 1 |
| Cation suppressor | 1 |
| Chromatography software | 1 |
| Column heater | 1 |
| PEEK flow path | 1 |
| Online gas-liquid separator | 1 |
| Spare parts kits | 1 |
| Vacuum degassing pump | 1 |
| Sand core filtration device | 1 |
| 0.22um syringe filter | 100 |
| Built-in anion eluent generator | 1 |
| Autosampler | 1 |
| Leakage Sensor | 1 |
| Gas-liquid separator | 1 |
Download catalog for
Ion chromatograph system LICS-A20 Catalog
Download manual for
Ion chromatograph system LICS-A20 Manual
Available Range :
Ion chromatograph system LICS-A20 : KOH/NaOH/LiOH/MSALabtron ion chromatography systems ensure excellent system stability and performance with capabilities for both anion and cation detection. They feature advanced conductivity detectors, auto-samplers, and intelligent workstations for stability and high accuracy. These systems include built-in eluent management, automatic cleaning modes and real time monitoring. Our ion chromatography systems have user-friendly interfaces and advanced technologies, ensuring reliability.



Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608