| Measuring range | 0 to 1200 ml |
| Measuring Accuracy | 0.02 |
| Display Accuracy | 1 ml |
| Liquid Proportion | 1.05 g/ml |
| Collection Speed | 0.5 ml/s to 3 ml/s |
| Collection Volume Range | 200 mL to 500 mL in 1 mL increments |
| Flow Rate Range | 10 to 150 m L/min |
| Swing Frequency | 30 rpm to 32 rpm |
| Swing Angle | 13 ° ± 2° |
| Alarm | Audio visual |
| Battery Life | 200 donations per full charge |
| Power | 12 W |
| Power Supply | AC 110 V / 220 V ± 10 %, 50 Hz / 60 Hz |
| External Dimension (W × D × H) | 218 × 172 × 160 mm |
| Weight | 2.8 kg |
Our Blood Collection Monitor LBMO-A10 is widely used in blood banks, hospitals, and mobile donation camps to ensure accurate, safe, and efficient blood collection.
Download catalog for
Blood Collection Monitor LBMO-A10 Catalog
Download manual for
Blood Collection Monitor LBMO-A10 Manual
Labtron provides a wide range of Blood Collection Monitor. They are user-friendly and designed for easy use in healthcare settings. They track blood flow in real time and provide immediate feedback to healthcare professionals. Their main function is to ensure precise blood collection. Our blood collection monitor enhances donor safety and the quality of blood products. Our blood collection monitor facilitates the gentle mixing of blood and anticoagulants, preventing clotting during collection. Renowned for superior quality, ensuring reliable performance and longevity.

The Blood Collection Monitor LBMO-A10 is an automated device designed to ensure safe and precise blood donation. It continuously monitors, mixes, and measures the collected blood, stopping the process automatically once the preset collection volume is achieved. This guarantees accurate volume control and improved donor safety.
Absolutely. The LBMO-A10 automatically halts blood collection upon reaching the preset volume, ensuring accurate and safe collection for both donor and operator.
The LBMO-A10 features a slip-free automatic clamp that activates when the programmed collection volume is reached, during pauses, or under low flow conditions. Additionally, it has a dedicated tube securement slot that keeps the tube firmly in place, eliminating the risk of accidental dislodgement.
The LBMO-A10 allows collection volumes from 200 mL to 500 mL, adjustable in 1 mL increments or better. This ensures precise volume customization for different donation requirements.
The LBMO-A10 supports a flow rate range of 10–150 mL/min, allowing flexible operation and smooth blood collection across varying donor conditions.
Post your question?
Our experts are here to help!
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom