Salt Spray Test Chamber LSSC-A12 is a nichrome heater comprised salt spray test chamber, designed as per ASTMB117, ISO9227 configurations, comes with 410 L of capacity and built-in 20 L salt solution tank with a brine mixing system. Features adjustable Atomizer tower and Spray nozzles spraying system with continuous/periodic spray, covered the entire working space to form a diffuse state. Equipped with PID controller, PT100 Ω / MV A-class sensor and timer, it offers complete recording of spray time with precise temperature and control. Designed with Glass fiber reinforced plastics housing, transparent resin lid and hydrostatic hood seal, it has sample holder of 6 round bars and 5 V-shaped grooves for different tests. With 31 L capacity of saturated air barrel protected against dry burning, this salt spray test chamber also offers protection against humidifier dry-combustion , over-temperature, over-current, water shortage and earth leakage.
Testing room capacity | 410 L |
Temperature range | Ambient to +60°C |
Humidity range | 95 to 98% RH |
Temperature Fluctuation | ± 0.5 ℃ |
Temperature Deviation | ± 2.0 ℃ |
Spray deposition | 1 to 2 ml / 80 cm2 .h |
Spray type | Continuous /Periodic |
Spraying system | Atomizer tower and Spray nozzle |
Air preheating | Saturated air barrel; 31 L |
Heating element | Nichrome heater |
Salt fog collection | Fog collector and Fog measure cylinder |
Standard sample holder | 8 round bars and 7 V-shaped grooves |
Controller | PID |
Sensor | PT100 Ω / MV A-class sensor |
Housing | Glass fiber reinforced plastics |
Lid | Transparent resin lid |
Sealing | Hydrostatic hood seal |
Protection |
Humidifier Dry-combustion Protection; Over-temperature Protection; Over-current Protection; Water Shortage Protection; Earth leakage Protection |
Maximum noise | 65 dBA |
Power supply | 220 V 50 Hz 1 P |
Water solution tank dimension (W×D×H) | 350×180×320 mm |
Internal dimension (W×D×H) | 1100×750×500 mm |
External dimension (W×D×H) | 1900×920×1200 mm |
Net weight | 260 kg |
Gross weight | 310 kg |
Salt spray test chamber is used to create and maintain the salt spray (fog) test environment, and test the anti-corrosion quality of all the materials surfaces after the rust-proof of painting, coating, electroplating, anodizing and rust-proof of greasing.
Model | Testing room capacity | Temperature range | Humidity range | Temperature Fluctuation | Temperature Deviation | Spray deposition | Spray type | Spraying system | Air preheating | Heating element | Salt fog collection | Standard sample holder | Controller | Sensor | Housing | Lid | Sealing | Protection | Maximum noise | Power supply | Water solution tank dimension (W×D×H) | Internal dimension (W×D×H) | External dimension (W×D×H) | Net weight | Gross weight |
LSSC-B10 | - | 10 to 90°C | 30 to 98% RH | - | - | - | - | - | - | - | - | - | Programmable color LCD touch screen controller, Ethernet connection, PC Link, USB | - | - | - | - | - | 65 dBA | 380 V, 50 Hz, 3 Phase | - | - | - | - | 657 kg |
LSSC-A10 | 110 L | Ambient to +60°C | 95 to 98% RH | ± 0.5 ℃ | ± 2.0 ℃ | 1 to 2 ml / 80 cm2 .h | Continuous /Periodic | Atomizer tower and Spray nozzle | Saturated air barrel; 31 L | Nichrome heater | Fog collector and Fog measure cylinder | 6 round bars and 5 V-shaped grooves | PID | PT100 Ω / MV A-class sensor | Glass fiber reinforced plastics | Transparent resin lid | Hydrostatic hood seal |
Humidifier Dry-combustion Protection; Over-temperature Protection; Over-current Protection; Water Shortage Protection; Earth leakage Protection |
65 dBA | 220 V 50 Hz 1 P | 350×180×320 mm | 600×460×400 mm | 1100×700×1000 mm | 110 kg | 185 kg |
LSSC-A11 | 326 L | Ambient to +60°C | 95 to 98% RH | ± 0.5 ℃ | ± 2.0 ℃ | 1 to 2 ml / 80 cm2 .h | Continuous /Periodic | Atomizer tower and Spray nozzle | Saturated air barrel; 31 L | Nichrome heater | Fog collector and Fog measure cylinder | 7 round bars and 6 V-shaped grooves | PID | PT100 Ω / MV A-class sensor | Glass fiber reinforced plastics | Transparent resin lid | Hydrostatic hood seal |
Humidifier Dry-combustion Protection; Over-temperature Protection; Over-current Protection; Water Shortage Protection; Earth leakage Protection |
65 dBA | 220 V 50 Hz 1 P | 350×180×320 mm | 1000×640×510 mm | 1580×910×1200 mm | 225 kg | 260 kg |
LSSC-A12 | 410 L | Ambient to +60°C | 95 to 98% RH | ± 0.5 ℃ | ± 2.0 ℃ | 1 to 2 ml / 80 cm2 .h | Continuous /Periodic | Atomizer tower and Spray nozzle | Saturated air barrel; 31 L | Nichrome heater | Fog collector and Fog measure cylinder | 8 round bars and 7 V-shaped grooves | PID | PT100 Ω / MV A-class sensor | Glass fiber reinforced plastics | Transparent resin lid | Hydrostatic hood seal |
Humidifier Dry-combustion Protection; Over-temperature Protection; Over-current Protection; Water Shortage Protection; Earth leakage Protection |
65 dBA | 220 V 50 Hz 1 P | 350×180×320 mm | 1100×750×500 mm | 1900×920×1200 mm | 260 kg | 310 kg |
LSSC-A13 | 780 L | Ambient to +60°C | 95% to 98% RH | ± 0.5 ℃ | ± 2.0 ℃ | 1 to 2 ml / 80 cm2 .h | Continuous /Periodic | Atomizer tower and Spray nozzle | Saturated air barrel; 31 L | Nichrome heater | Fog collector and Fog measure cylinder | 8 round bars and 7 V-shaped grooves | PID | PT100 Ω / MV A-class sensor | Glass fiber reinforced plastics | - | Pneumatic seal |
Humidifier Dry-combustion Protection; Over-temperature Protection; Over-current Protection; Water Shortage Protection; Earth leakage Protection |
65 dBA | 220 V 50 Hz 1 P | 600×370×350 mm | 1300×1000×600 mm | 2200×1200×1400 mm | 280 kg | 450 kg |
LSSC-A14 | 1030 L | Ambient to +60°C | 95% to 98% RH | ± 0.5 ℃ | ± 2.0 ℃ | 1 to 2 ml / 80 cm2 .h | Continuous /Periodic | Atomizer tower and Spray nozzle | Saturated air barrel; 31 L | Nichrome heater | Fog collector and Fog measure cylinder | 8 round bars and 7 V-shaped grooves | PID | PT100 Ω / MV A-class sensor | Glass fiber reinforced plastics | - | Pneumatic seal |
Humidifier Dry-combustion Protection; Over-temperature Protection; Over-current Protection; Water Shortage Protection; Earth leakage Protection |
65 dBA | 220 V 50 Hz 1 P | 600×370×350 mm | 1600×900×720 mm | 2500×1300×1600 mm | 390 kg | 560 kg |
LSSC-A15 | 1600 L | Ambient to +60°C | 95% to 98% RH | ± 0.5 ℃ | ± 2.0 ℃ | 1 to 2 ml / 80 cm2 .h | Continuous /Periodic | Atomizer tower and Spray nozzle | Saturated air barrel; 31 L | Nichrome heater | Fog collector and Fog measure cylinder | 8 round bars and 7 V-shaped grooves | PID | PT100 Ω / MV A-class sensor | Glass fiber reinforced plastics | - | Pneumatic seal |
Humidifier Dry-combustion Protection; Over-temperature Protection; Over-current Protection; Water Shortage Protection; Earth leakage Protection |
65 dBA | 220 V 50 Hz 1 P | 600×370×350 mm | 2000×1000×800 mm | 3000×1420×1850 mm | 480 kg | 630 kg |