| Measurement Range | 0 to 100 PCU |
| Minimum Resolution | 0.01PCU |
| Indication Error | Less than or equal to 3% |
| Repeatability | Less than or equal to 2% |
| Hourly Drift | Less than or equal to 0.1PCU |
| Working temperature | 5° to 40°C |
| Relative humidity | Less than or equal to 85% |
| Power Supply | AC 220 (165V-240V) |
| Dimension | 387mm × 292mm × 134mm |
| Weight | 2.6kg |
Our Benchtop Colorimeter LCOM-A13 with its high quality and multiple measurement modes is widely applicable to various industries for the measurement of drinking water, surface water, groundwater and other water samples.
| Accessory | Quantity |
| Main Unit | 1 pc |
| Cuvette | 2 pc |
| Power Cord | 1 pc |
Download catalog for
Benchtop Colorimeter LCOM-A13 Catalog
Labtron Benchtop Colorimeters have a wide range of light filters to ensure accurate concentration and color measurements for diverse samples. These Colorimeters utilize high-quality silicon photodiodes for reliable detection of light. They offer dual knobs for precise focusing on samples of different opacity for accurate readings. Our Colorimeters have durable constructions, LED displays, efficient optical systems and user-friendly design.




Benchtop Colorimeter measures the water samples using the Platinum-Cobalt Unit (PCU) scale, which is commonly used in water quality testing. It has a measuring range of 0 to 100 PCU with a minimum resolution of 0.01 PCU. It features an indication error of less than or equal to 3% and repeatability of less than or equal to 2%, ensuring high precision and reliable results.
The Benchtop Colorimeter comes with a calibration function. It allows for self-calibration during operation, enhancing accuracy and reducing downtime. Curve queries are possible, including not only the curve equations but also the standard values of each standard substance used during calibration, corresponding absorbance values, calibration time, and calibration personnel ID fully retrieving the calibration state.
The Benchtop Colorimeter uses a cold light source and a narrow-band interference optical system, which ensures excellent optical stability and low drift (less than or equal to 0.1 PCU per hour).
The Benchtop Colorimeter is designed to operate within a temperature range of 5°C to 40°C and relative humidity of less than or equal to 85%. It requires a power supply of AC 220V (165V–240V). For optimal performance, it should be used in an environment free from significant vibration, electromagnetic interference, and direct sunlight exposure.
Benchtop Colorimeter allows for self-calibration during operation, enhancing accuracy and reducing downtime. Supports curve queries that retrieve both the curve equations and the full calibration conditions. It includes a USB interface that allows connection to a computer and supports single-item or page printing of test records, making it convenient for documentation and reporting.
Benchtop Colorimeter measures the water samples using the Platinum-Cobalt Unit (PCU) scale, which is commonly used in water quality testing. It has a measuring range of 0 to 100 PCU with a minimum resolution of 0.01 PCU. It features an indication error of less than or equal to 3% and repeatability of less than or equal to 2%, ensuring high precision and reliable results.
The Benchtop Colorimeter comes with a calibration function. It allows for self-calibration during operation, enhancing accuracy and reducing downtime. Curve queries are possible, including not only the curve equations but also the standard values of each standard substance used during calibration, corresponding absorbance values, calibration time, and calibration personnel ID fully retrieving the calibration state.
The Benchtop Colorimeter is designed to operate within a temperature range of 5°C to 40°C and relative humidity of ?85%. It requires a power supply of AC 220V (165V–240V). For optimal performance, it should be used in an environment free from significant vibration, electromagnetic interference, and direct sunlight exposure.
Benchtop Colorimeter allows for self-calibration during operation, enhancing accuracy and reducing downtime. Supports curve queries that retrieve both the curve equations and the full calibration conditions. It includes a USB interface that allows connection to a computer and supports single-item or page printing of test records, making it convenient for documentation and reporting.
The Benchtop Colorimeter uses a cold light source and a narrow-band interference optical system, which ensures excellent optical stability and low drift (less than or equal to 0.1 PCU per hour).
Post your question?
Our experts are here to help!
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom