| Measuring range | Temperature: -5.0°C to 105.0°C pH: 0.00 to 14.00 mV: -1800 to 1800 |
| Resolution | Temperature: 0.1°C pH: 0.01 mV: 0.1 |
| Basic error | Temperature: ±0.3°C ± 1 word pH: ±0.01 ± 1 word mV: ±0.03% FS |
| Control temperature | 60°C ±1°C |
| Burette capacity tolerance | 10 ml burette: ± 0.025 ml 20 ml burette: ±0.035 ml |
| Burette resolution | 10 ml burette: 1/10000 20 ml burette: 2/10000 |
| Burette infusion speed | 50 ± 10 s (when burette is full) |
| Titration analysis repeatability | 0.002 |
| Titration control sensitivity | ± 2 mV |
| Stability | (± 0.3 mV ± 1 word)/3 h |
| Temperature controller | PID |
| Display | Digital |
| Heating tube | Electric |
| Power supply | AC 220 ± 22 V, 50 ± 1 Hz |
| Host apparatus dimension (W×D×H) | 460×480×580 mm |
| Host apparatus weight | 36 kg |
| Titration device dimension (W×D×H) | 460×700×1140 mm |
| Titration device weight | 55 kg |
Automatic Crude Oil Salt Content Analyzer is used for determination of chloride (salt) content in crude oils across refineries, pipelines, terminals, laboratories, inspection companies, oil platforms etc.
Labtron Crude Oil Salt Content Analyzer features a microcomputer-controlled unit that ensures precise control and efficiency, allowing for highly accurate titration results. These are innovative titration cells with a sand core that enhances durability. Their electric heating tube with adjustable voltage heating users controls the temperature, ensuring optimal conditions for the reaction. They automatically determine the titration endpoint through the potentiometric titration method, ensuring accuracy by detecting the exact point. Our crude oil salt content analyzer features a standard titration extraction device that simplifies sample preparation.

Our Automatic Crude Oil Salt Content Analyzer LCOS-A10 uses the potentiometric titration method to check the salt level in crude oil by tracking voltage changes during the test. This helps find the exact point where the salt reacts, even in dark or unclear samples where visual checks don’t work. It gives steady and correct results with all types of crude oil. This makes it a useful tool for regular salt testing in petroleum labs.
Labtron’s Automatic Crude Oil Salt Content Analyzer LCOS-A10 has adjustable burette and titration settings to handle different sample sizes based on testing needs. It works well with light, medium, and heavy crude oils, making it useful for many types of labs. The system controls the flow and mixing speed carefully to give steady and correct results with different oil types. This makes it a good choice for labs that test many kinds of samples.
The Automatic Crude Oil Salt Content Analyzer LCOS-A10 features an automatic cleaning system that flushes the titration path after each analysis, minimizing cross-contamination between samples. It includes functions like auto-replenishment and fixed-volume liquid addition, reducing the manual steps needed for preparation. This automation ensures consistency in each test cycle, especially in high-volume labs. The built-in procedures also enhance safety by limiting operator exposure to heated oil or reagents.
Our Automatic Crude Oil Salt Content Analyzer LCOS-A10 is built for continuous use, making it suitable for labs that handle many samples daily. It has automatic features like self-cleaning, preset test cycles, and accurate control of reagents to run tests one after another with little delay. The system uses strong parts and stable controls to keep working without problems. With regular care and simple checks, it can run smoothly all day.
Labtron’s Automatic Crude Oil Salt Content Analyzer LCOS-A10 includes controlled heating with PID regulation, limiting temperature fluctuations and preventing overheating. Sample handling is enclosed, and the heating tubes operate within a safe, preset range to reduce risks to users. Because most steps are automated, direct contact with hot surfaces or chemicals is minimized. This combination of design safety and automation enhances both operator protection and lab compliance.
The Automatic Crude Oil Salt Content Analyzer LCOS-A10 is designed for low-maintenance operation, thanks to its automated cleaning system and stable digital controls. While periodic calibration is necessary to maintain accuracy, it can be done easily using standard solutions and internal software prompts. Its stable titration delivery and precise temperature control minimize drift and the need for frequent adjustments. Routine maintenance is simple and does not require specialized tools or extensive downtime.
Post your question?
Our experts are here to help!
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom