| Operation Pressure | 100 to 10000 mbar (150 psi) |
| Operation Condition | Ambient Temperature: +5℃ to +40℃ Relative Humidity: 10 to 80% |
| Test Accuracy | Upstream Volume Test: ±4% Bubble Point Test: ±50mbar Diffusion Flow Test: ±4% Water Immersion Test: ±0.01ml |
| Test Scope | BP: 100 to 8000mbar DF: 1 to 900ml/min WI: 0.01 to 100ml/min |
| Audit Trail | Audit trail records are exportable and irreversible |
| Audit Trail Record Storage | More than or equal to 5 years storage |
| Reservation | Solution (programs): 1000 sets |
| Dust and Splash Level | IP54 |
| Print Function | Built-in printer, can control on PC, subject selection available |
| Historical Record Storage | 300000 sets |
| Operating System | Linux system |
| Display Screen | High-definition 10-inch color touch screen |
| Serial Connection | RS232/USB/Support wireless connection |
| Applicable Environment | Above D level |
| Power Consumption | 120W |
| Power Supply | 100 to 240 VAC, 50HZ |
| Dimension (W×D×H) | 300mm × 480mm × 210mm |
| Packaging Dimension | 710mm × 470mm × 600 mm |
| Weight | 10 kg |
| Gross Weight | 27 kg |
High-accuracy pressure sensor integration
Automatic self-test with fault reporting
Stores 1000 pre-set programs
Display curve for diffusion flow
Optimized air inlet control unit
Supports USB data export function
Expandable communication and network interface
Wireless communication functions available
Staubli connectors prevent incorrect connections
Extended gas path up to 100m
Built-in thermal printer included
Multiple switchable pressure units available
Tests up to 12-core filters
Customized Anti-backflow device available
Our Filter Integrity Tester LIT-A10 ensures precise absorbance measurement, spectral scanning, and quantitative analysis required for symmetric and asymmetric membrane test, needle filters, capsule filters, flat filters, cartridge filters, ultrafiltration membrane packages, columns and various irregular filters making it ideal for research applications.
Download catalog for
Filter Integrity Tester LIT-A10 Catalog
Labtron Filter Integrity Tester utilize correct pressure regulators for maintaining consistent and accurate testing conditions. They feature user-friendly touchscreen controls for easy navigation and streamlined operations. Their high-sensitivity sensors detect minute leaks, ensuring compliance with stringent industry standards. They offer automated data logging, assuring accurate and tamper-proof record-keeping of test results. Our Filter Integrity Tester provide reproducible results, making them ideal for pharmaceutical, biotech, and food industries.

Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608