Light source | 150 W Xenon lamp |
Excitation wavelength | 200 nm to 900 nm |
Emission wavelength | 200 nm to 900 nm |
Excitation slit | 2 nm, 5 nm, 10 nm, 20 nm |
Emission slit | 2 nm, 5 nm, 10 nm, 20 nm |
Wavelength accuracy | ±1.0 nm |
Wavelength repeatability | ≤ 0.5 nm |
Signal-to-Noise Ratio | Raman peak of water (P-P): S/N ≥ 150 (10 nm Slit) |
Detection limit | ≤1×10-10 g/ml (Quinine Sulphate Solution) |
Linearity | γ ≥ 0.995 |
Peak repeatability | ≤ 1.5% |
Stability (10 min) Zero Drift | ±0.3 |
Stability (10 min) Value Limit | ±1.5% |
Wavelength scan speed | Multi-speed level, Maximum at 48000 nm/min |
Photometric value | 0.00 to 10000.00 |
Data interface | USB 2.0 |
Maximum power consumption | 200 W |
Power Source | AC 220 V/50 Hz; 110 V/60 Hz |
Dimension | 380×445×310 mm |
Net weight | 12 kg |
Gross weight | 14 kg |
Accessories No. | Accessories Name | Functions |
1 | Membrane sample rack | For membrane samples |
2 | Powder sample rack | For powder samples |
3 | Up conversion fluorescent holder | Base holder for other racks |
4 | Octave Double-frequency elimination filter (300 nm,350 nm,400 nm,450 nm,500 nm,550 nm,600 nm,650 nm for one set) | Remove the interference of double-frequency peaks of these wavelengths |
5 | Auto polarization filter | - |
6 | Sample cell jacket | - |
7 | Peltier temperature (15 to 40°C) control holder | - |
Download catalog for
Fluorescence Spectrophotometer LFS-A12 Catalog
Available Range :
Fluorescence Spectrophotometer LFS-A10 : 150 W Xenon lampSimilar Products
Address:Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 20 8004 3587 Phone: +44 2080 043608