| Residual liquid | ≤ 1 µL |
| Washing heads | 1 × 8 or 1 × 12 heads |
| Plate types | Flat , U , V , C bottom |
| Washing method | Sling-point, two-point and multi-point |
| Washing volume | 50 to 2000 µL, increase by 50 µL |
| Washing times | 1 to 99 times |
| Washing channels | 3 |
| Bottles | 3pcs 2.5L washing bottle (1pcs, 2.5 L washing bottle) |
| Dispense accuracy | ≤ 2% @ 300 µL |
| Liquid injection accuracy | CV ≤ 1.5 % @ 300 µL |
| Soaking and shaking time | (99’00”) |
| Display | 4.3-inch color LCD display |
| Program Storage Capacity | Up to 100 programs |
| Date connection | USB |
| Power supply | AC 110 to 240 V, 50/60 Hz |
| Weight | 12 kg |
Our Microplate Washer LMPW-A10 is suitable for washing microplates of various specifications, including flat-bottom, U-shaped, V-shaped, and C-shaped plates.
Download catalog for
Microplate Washer LMPW-A10 Catalog
Labtron Microplate Washer comes with a wash, rinse, buffer, and waste bottles for washing needs. It is adaptable to different microplate formats. These can store ample programs for customized washing protocols. They facilitate ease of use with intuitive navigation and operation. Our Microplate Washer maintains cleanliness with adjustable flushing times. Alerts users to prevent liquid depletion or overflow, enhancing safety and efficiency. They maintain consistent performance through automated vacuum and pressure monitoring.







The Microplate Washer LMPW-A10 offers single-point, two-point, and multi-point washing options, which improve the efficiency of different assays. These modes allow you to adjust the wash coverage based on plate design or assay sensitivity. Multi-point mode ensures thorough cleaning, especially for wells with high protein binding. This versatility supports better assay optimization and cleaner backgrounds.
Labtron’s Microplate Washer LMPW-A10 includes features like a built-in filter, precise aspiration, and low residual volume to reduce the risk of cross-contamination. It allows for programmable washing sequences to ensure thorough cleaning between cycles. By automating fluid handling, it also minimizes human contact, which further reduces contamination chances. It’s especially helpful in assays involving infectious agents or low-concentration samples.
Our Microplate Washer LMPW-A10 ensures thorough removal of unbound reagents, directly enhancing the accuracy and sensitivity of ELISA results. Insufficient or uneven washing can cause high background signals, false positives, and poor reproducibility. It delivers uniform wash performance across all wells, reducing variability and improving assay reliability. In both diagnostic and research applications, this level of consistency is critical for generating accurate and dependable results.
The Microplate Washer LMPW-A10 offers soak and shake functions that allow reagents to rest in the wells for a set time and gently agitate them during the wash cycle. This helps improve the removal of unbound reagents or cells that may adhere to the well surface. It enhances overall washing efficiency in protocols requiring thorough yet gentle cleaning. These functions are particularly beneficial in sensitive applications like cell-based or bead-based assays.
Our Microplate Washer LMPW-A10 features a built-in pipeline flushing function designed to clean internal fluid lines using distilled water. This prevents salt, buffer, or protein buildup that could clog the system or compromise washing performance. Regular flushing helps maintain consistent results and reduces the need for frequent manual maintenance. It also extends the instrument’s lifespan, especially when used with varied reagents or after heavy workloads.
Post your question?
Our experts are here to help!
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608