| DO Measuring Range | 0.00 to 20.0 mg/L; 0.0 to 200% |
| Temperature Measuring Range | 0 to 60℃ (ATC/MTC) |
| Atmosphere Measuring Range | 300 to 1100 hPa |
| DO Resolution | 0.01 mg/L, 0.1 mg/L (ATC); 0.1% / 1% (ATC) |
| Temperature | 0.1℃ |
| Atmosphere | 1 hPa |
| Electronic Unit Measurement Error DO | ±0.5 % FS |
| Electronic Unit Measurement Error Temperature | ±0.2℃ |
| Electronic Unit Measurement Error Atmosphere | ±5 hPa |
| Calibration | Up to 2 points (water vapor saturated air / zero oxygen solution) |
| Display | LCD |
| Sensor Input Connector | BNC |
| Data Storage | Calibration data; 99 groups of measurement data |
| Working Condition Temperature | 5 to 40℃ |
| Relative Humidity | 5% to 80% (non-condensing) |
| Installation Grade | Ⅱ |
| Pollution Grade | 2 |
| Altitude | ≤2000 m |
| Power Supply | DC 6V / 20 mA; 4 × AA/LR6 1.5 V or NiMH 1.2 V (rechargeable) |
| Dimension | 230 × 100 × 35 mm |
| Net Weight | 0.4 Kg |
| Gross Weight | 2 Kg |
LCD
Data storage
Calibration
One-hand operation
Low power consumption
Portable Dissolved Oxygen Analyzer LPDOA-A10 is ideal for on-site water analysis, wastewater treatment, environmental monitoring and research applications across a wide range of industries.
Labtron’s Portable Dissolved Oxygen Analyzer are compact and portable devices for analysis of dissolved oxygen in different water samples. They are equipped with an LCD display that offers dependable and easy operation, even under low-light outdoor conditions. They include temperature compensation system, which automatically adjusts readings according to the samples temperature. Offers excellent anti-corrosion resistance, along with dustproof and waterproof protection. Our Portable Dissolved Oxygen Analyzer are ideal for aquaculture, laboratory testing and research units across various industries.

The Labtron LPDOA-A10 is designed as a lightweight, compact, and easy-to-carry device, making it perfect for on-site applications. It comes with one-hand operation, low power consumption, and runs on AA or rechargeable batteries, allowing long hours of field testing without frequent recharging. Its rugged design ensures reliable performance even in outdoor environments such as rivers, lakes, and wastewater treatment sites.
The Labtron LPDOA-A10 provides precise readings with a measuring range of 0.00 to 20.0 mg/L and 0.0 to 200% saturation. It has a high-resolution of 0.01/0.1 mg/L or 0.1%/1% (ATC) and an error margin of only ±0.5% FS. This makes it highly reliable for detecting both very low and very high oxygen levels in water samples, which is essential for accurate water quality monitoring in laboratories, environmental studies, and industrial processes.
The Labtron LPDOA-A10 supports up to two-point calibration, which includes calibration with water vapor-saturated air and a zero-oxygen solution. This ensures that the analyzer delivers stable and accurate readings under varying environmental conditions. With this calibration system, users can trust the analyzer for both routine and advanced testing requirements.
Yes, the Labtron LPDOA-A10 is equipped with data storage capability. It can store calibration data along with up to 99 groups of measurement data. This is particularly useful for field researchers and environmental scientists who need to collect multiple samples in different locations. Stored data can later be reviewed, reducing the need for manual note-taking and ensuring accuracy in record-keeping.
The Labtron LPDOA-A10 is a versatile analyzer widely used across industries that require monitoring of dissolved oxygen. It is ideal for on-site water analysis, wastewater treatment plants, aquaculture, environmental monitoring, and laboratory research. By providing accurate readings, it helps ensure water quality, environmental safety, and compliance with regulatory standards.
Post your question?
Our experts are here to help!
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608