Sand and Dust Test Chamber LSC-A12 operates within a temperature range from ambient to +50°C and relative humidity of less than 30% RH. The chamber employs a circulatory pump system that recycles dust for consistent testing. An enclosed heater ensures the dust remains dry throughout the process. Designed to meet the CE and ISO standards, our test chamber offers safety features for current leakage, overload, over current, short circuit, over temperature, motor heat, and pressure.
Request QuoteTemperature range | Ambient to + 50 ℃ |
Relative Humidity | < 30 % RH |
Normal wire diameter | 50 µm |
Normal width of a gap between wire | 75 µm |
The Talcum powder amount | 2 kg to 4 kg / m3 |
Fight time | 0 to 99 H 59 M |
Fan cycle time | 0 to 99 H 59 M |
Specimen power outlet | Dust-proof socket AC220V 16A |
Vacuum system | Equipped with a pressure gauge, air filter, pressure regulator, connecting tube |
Circulating fan | Enclosed alloy low-noise motor, multi-vane centrifugal fan |
Heating system | Independent Nichrome electronic heating system |
Power Consumption | 7500 W |
Power supply | 380 V 50 Hz |
Internal dimensions | 1000 × 1500 × 1000 mm |
Overall dimensions | 1250 × 1900 × 1960 mm |
Sand and Dust Test Chamber ensures that equipment remains reliable and effective when exposed to sand and dust in real-world scenarios. Used to test the quality of products such as electronics, plastic products, electrical appliances, metals, chemicals, building materials, aerospace, medical care, food, and beverages.