Disintegration tester LDIT-A12 is adopted with high-precision digital electronic sensor, ensures quality precision and accuracy of the water bath temperature without need for calibration. Equipped with two to three sets of basket, conducts independent disintegration experiments. The default setting for lifting time of the basket is 15 mins, can manually set the other time value as per the experimental need.
Request QuoteNumber of basket lifting set | 3 sets |
Lifting frequency of the basket | 30 to 32/ min |
Lifting distance of basket | 55 ± 1 mm |
Minimum spacing from screen to bottom of beaker | 25 ± 2 mm |
Mesh size | 2 mm (as per standard configuration ) |
Automatic time range | 5 to 120 min |
Work noise | < 60 dB |
Temperature prefabrication range | Room Temperature to 50°C, display resolution: 0.1 °C |
Temperature control accuracy | ± 0.3°C |
Time prefabrication range | 1 to 900 min |
Time control accuracy | 0.5 min |
Accuracy of temperature control | 37 ± 0.5°C |
Dimension (L × W × H) | 540 × 320 × 440 mm |
Power supply | 220 V/50 Hz |
Total power | 600 W |
Gross Weight | 25 kg |
Used for detection of rate of disintegration of solids such as tablets, capsules, pills in required /set (prescriptive) conditions at pharmaceutical industry.
Accessories no. | Name | Capacity | Qty. |
1 | Beaker | 1000 ml | 3 |
Accessories no. | Name |
1 | 0.425 mm diameter mesh |
2 | 1 mm diameter mesh |
3 | Beaker |
Available Range :
Number of basket lifting set : 1 set Number of basket lifting set : 2 sets Number of basket lifting set : 3 setsRelated Models :
Model | Number of basket lifting set | Lifting frequency of the basket | Lifting distance of basket | Minimum spacing from screen to bottom of beaker | Mesh size | Automatic time range | Work noise | Temperature prefabrication range | Temperature control accuracy | Time prefabrication range | Time control accuracy | Accuracy of temperature control | Dimension (L × W × H) | Power supply | Total power | Gross Weight |
LDIT-A10 | 1 set | 30 to 32/ min | 55 ± 1 mm | 25 ± 2 mm | 2 mm (as per standard configuration ) | 5 to 120 min | < 60 dB | Room Temperature to 50°C, display resolution: 0.1 °C | ± 0.3°C | 1 to 900 min | 0.5 min | 37 ± 0.5°C | 260 × 300 × 440 mm | 220 V/50 Hz | 600 W | 15 kg |
LDIT-A11 | 2 sets | 30 to 32/ min | 55 ± 1 mm | 25 ± 2 mm | 2 mm (as per standard configuration ) | 5 to 120 min | < 60 dB | Room Temperature to 50°C, display resolution: 0.1 °C | ± 0.3°C | 1 to 900 min | 0.5 min | 37 ± 0.5°C | 400 × 320 × 440 mm | 220 V/50 Hz | 600 W | 20 kg |
LDIT-A12 | 3 sets | 30 to 32/ min | 55 ± 1 mm | 25 ± 2 mm | 2 mm (as per standard configuration ) | 5 to 120 min | < 60 dB | Room Temperature to 50°C, display resolution: 0.1 °C | ± 0.3°C | 1 to 900 min | 0.5 min | 37 ± 0.5°C | 540 × 320 × 440 mm | 220 V/50 Hz | 600 W | 25 kg |