| Indication error | Less than or equal to ±5% |
| Repeatability | Less than or equal to 2% |
| Optical stability | Absorbance drift less than or equal to 0.002A/20min |
| Working temperature | 5° to 40°C |
| Relative humidity | Less than or equal to 85% |
| Power Consumption | 30W |
| Power Supply | AC 220 (165V-240V) |
| Dimension | 387 × 292 × 134 mm |
| Packaging Dimensions | 900 × 540 × 640 mm |
| Weight | 2.6 Kg |
| Gross Weight | 10.5 Kg |
Our Heavy Metal Analyzer LHVMA-A10 is well-suited for determination of freshwater, surface water, sewage and industrial wastewater.
Download catalog for
Heavy Metal Analyzer LHVMA-A10 Catalog
The Heavy Metal Analyzer allows flexible parameter combinations, enabling detection of various heavy metals such as copper, nickel, and more based on user requirements. Offers high accuracy with an indication error of less than or equal to 5% and repeatability of less than or equal to 2%. Ideal for freshwater, surface water, sewage, and industrial wastewater analysis.
The Heavy Metal Analyzer stores up to 200 curves and 3000 records and save 100,000 data records. It ensures data protection even when power-off. During record queries, single-item or page printing can be performed. Curve queries can include not only curve equations but also the standard values of each standard substance used during calibration.
The Heavy Metal Analyzer is designed to operate within a temperature range of 5°C to 40°C and relative humidity of less than or equal to 85%. It requires a power supply of AC 220V (165V–240V). For optimal performance, it should be used in an environment free from significant vibration, electromagnetic interference, and direct sunlight exposure.
The Heavy Metal Analyzer allows flexible parameter combinations, enabling detection of various heavy metals such as copper, nickel, and more based on user requirements. Allows users to customize the wavelength anywhere within the 420nm to 700nm range for specific detection needs and ensures excellent optical stability, with an absorbance drift less than or equal to 0.002A/20 minutes. Ideal for analysis of freshwater, surface water, sewage, and industrial wastewater in labs, environmental monitoring units, and treatment facilities.