Request A Quote
Oxidation Stability Tester LOST-A10

Oxidation Stability Tester LOST-A10

Oxidation Stability Tester LOST-A10 maintains insulating oil testing conditions with a PID controller, ensuring 200°C ±0.1°C stability. It keeps insulating oil at 120°C while exposing it to a copper catalyst and controlled airflow. The unit has four positions with flow meters regulating airflow from 0 to 120 mL/min. Its stainless steel structure has a ceramic coating, enhancing durability and resistance to high temperatures. Ideal for assessing oxidation stability in petroleum products, insulating oils, and hydrocarbons.

Request Quote
Heating mode Metal bath
Temperature range Ambient to 200°C ±0.1°C
Temperature control PID controller
Working units 4 pieces
Flow control 0 to 120 mL/min
Air filtration Glass wool filter
Purification system NaOH bottle, H₂SO₄ washing bottle, safety bottle
Standards SH/T0811, IEC61125
Power supply AC 220V, 50Hz
Total power 2.5 kW
Dimensions (L×W×H) 430 × 540 × 630 mm
Net weight 30 kg
Gross weight 48 kg
  • Microprocessor-controlled temperature regulation
  • Efficient thermal insulation ensures stability

  • Integrated overheating protection system

  • Advanced three-stage filtration

  • Intuitive user-friendly interface

  • Efficient multi-sample capacity

  • Robust construction material

Our Oxidation Stability Tester LOST-A10 evaluates oxidation stability of insulating liquids in transformers, power plants, and labs. It helps assess oil degradation, ensuring reliable performance in insulation oil quality analysis.

Accessory Name

Quantity

Silicone Tube Φ4

2 pcs

Fuse

4 pcs

Temperature Sensor

1 pc

Sand Paper 220

5 pcs

Air Pipe Φ6

3 meters

Accessory Name

Quantity

Oxidation Tube

4 pcs

Absorption Tube

4 pcs

Glass Filter

1 set

Soap Bubble Flow Meter

1 set

Catalytic Copper Wire

50 pcs

Models of Oxidation Stability Tester

Labtron Oxidation Stability Tester assess resistance of petroleum products, lubricants, and fuels to oxidative degradation at temperature upto 380℃. They utilize thermal-controlled chamber to maintain consistent testing conditions for samples. Their programmable timers allow for accurate control over duration of each test cycle. They incorporate pressure sensors to monitor changes in the sample during oxidation. Our Oxidation Stability Tester is applicable for product development, storage and stability studies and comparative analysis.

Oxidation Stability Tester LOST-D20

Oxidation Stability Tester LOST-D20

  • Temperature Range : Ambient to 150℃
  • Temperature Precision : ± 0.1°C
  • Test Unit : 2 units
  • Ambient Temperature : (Room temperature to 35°C
Oxidation Stability Tester LOST-E10

Oxidation Stability Tester LOST-E10

  • Temperature Range : 100℃ to 380℃
  • Temperature Accuracy : ± 1℃
  • Pressure Range : 0 to 340 mmHg
  • Pressure System : 500psi
Automated Oxidation Stability Tester LOST-F10

Automated Oxidation Stability Tester LOST-F10

  • Temperature Control : 150.0°C ± 0.1°C; 140.0°C ± 0.1°C
  • Sample Size : 50g ± 0.5g
  • Wire Catalyst : Electrolytic copper wire conforming to ASTM D2272, D2112
  • Vessel Angle : 30°
Biodiesel Oxidation Stability Tester LOST-D14
Biodiesel Oxidation Stability Tester LOST-D14
  • Temperature Range : RT to 150 ℃
  • Temperature Accuracy : ±0.1℃
  • Temperature Control : PID digital display temperature controller
  • Sample Volume : 5 g
Automatic Lubricating Oils Oxidation Stability Tester LOST-D10
Automatic Lubricating Oils Oxidation Stability Tester LOST-D10
  • Samples : 2
  • Oil bath cubage : 30 L
  • Angle between bomb and water surface : 30 °
  • Speed of rotary assembly : 100 ± 5 RPM
Transformer Oil Oxidation Stability Tester LOST-D11

Transformer Oil Oxidation Stability Tester LOST-D11

  • Samples : 6
  • Temperature Controlling Range : Room temperature to 160℃
  • Temperature Controlling Accuracy : ± 0.2℃
  • Temperature Controlling Mode : Automatic digital temperature controller
Oxidation Stability Tester LOST-A11

Oxidation Stability Tester LOST-A11

  • Flow Range : 8 to 12 L/h ±10%
  • Temperature Controlling : Import PID temperature controller
  • Measuring Range : 0 to 800 μS/cm
  • Resolution : 0.1 μS/cm
Distillate Fuel Oil Oxidation Stability Tester LOST-D12

Distillate Fuel Oil Oxidation Stability Tester LOST-D12

  • Samples : 8
  • Temperature : 100 ℃
  • Temperature controlling accuracy : 0.1 ℃
  • Flow meter : 3 ± 0.3 L / H oxygen