| Capacity | 18 L |
| Working Sterilization Temperature | 126℃ |
| Working Sterilization Pressure | 0.142 Mpa |
| Design Pressure | 0.165 Mpa |
| Design Temperature | 129℃ |
| Voltage | 220 V |
| Power Consumption | 2 Kw |
| Sterilization Room Dimensions | Ø280 × 235 mm |
| Packing Dimension (L × W × H) | 40.5 × 40.5 × 43 cm |
| Gross Weight | 12.5 kg |
| Net Weight | 14.2 kg |
Overpressure Self-Discharge System
High Temperature Tolerance
Dual Scale Pressure Gauge
Self-Expanding Seal Technology
Durable Stainless Steel Build
Portable Autoclave LPOA-C10 is a lightweight, mobile sterilization device that uses steam under pressure to disinfect medical and laboratory equipment. It is ideal for use in fieldwork, small clinics, and remote locations for convenient and effective sterilization.
Accessories Name | Quantity |
Dual Scale Pressure Gauge | 1 |
Self-Expanding Sealing Ring | 1 |
Anti-dry Heater | 1 |
Automatic Temperature Control Safety Valve | 1 |
Download catalog for
Portable Autoclave LPOA-C10 Catalog
Labtron Portable Autoclave provides efficient sterilization at temperature of 126℃ in compact environment. They are designed for easy transportation. Their user-friendly controls and minimal setup simplify the disinfection process and ensure ease of use. They are equipped with safety mechanism to prevent accidents. Our Autoclave is convenient solution in small research facilities, veterinary practices, field clinics and rural healthcare.



.webp)

.webp)
.webp)

Labtron portable autoclaves are specifically designed for applications where mobility, compactness, and efficiency are paramount. These units are ideal for small medical clinics, dental practices, veterinary settings, mobile labs, and on-field medical camps, especially in rural or emergency areas where a traditional autoclave setup isn’t feasible. Their lightweight and compact structure ensures they can be easily carried, stored, and set up without requiring complex infrastructure. Despite their size, Labtron ensures these autoclaves maintain consistent steam pressure and temperature for effective sterilization. This makes them reliable for sterilizing essential items like forceps, syringes, scissors, and glass containers during medical interventions or scientific testing on-site.
Labtron portable autoclaves are capable of sterilizing a broad range of heat-resistant medical and laboratory items, including surgical instruments, glassware, metal tools, cotton swabs, gauze, and certain types of rubber and plastic equipment. These autoclaves utilize saturated steam under pressure typically at 121°C and 15 psi to effectively eliminate all microbial life, including bacteria, viruses, fungi, and spores. However, they are not suitable for sealed containers, flammable liquids, or sensitive electronics, which could be damaged under pressure and high heat. Labtron includes user manuals with clear loading instructions and material compatibility charts to help users avoid damage and ensure the most effective sterilization process.
Labtron offers both electrically powered and non-electric models, such as kerosene, LPG, or spirit-heated units, to cater to a variety of environments. For laboratories and clinics with regular electricity, the electric models are plug-and-play, offering automated cycles and precise control. In contrast, for remote areas, mobile medical teams, or disaster response units, the non-electric variants offer complete independence from the power grid. These are particularly useful in developing regions or emergency situations, where reliable sterilization is needed despite lack of infrastructure. Labtron designs both models with easy-to-use heating systems, robust chambers, and durable materials to withstand repeated transportation and usage.
Labtron ensures that even its portable autoclaves include reliable pressure gauges and temperature indicators. These allow users to keep a real-time check on internal conditions and ensure the sterilization temperature (typically 121°C) and pressure (around 15 psi) are achieved and maintained for the required duration generally 15 to 30 minutes, depending on the load. Some advanced models may include alarm features that indicate when sterilization is complete or if unsafe pressure levels are approached. All units are fitted with safety release valves that automatically expel excess steam, preventing accidents. This user-friendly interface allows healthcare workers and lab technicians to confidently operate the autoclave, even with minimal technical training.
Labtron emphasizes regular maintenance to ensure optimal performance and safety. After every use, users should drain the chamber, remove residual water, and wipe down the interior to prevent rust or scale deposits. Weekly maintenance should include checking the gasket, pressure valve, and safety mechanisms for wear and tear. The heating element or burner depending on whether it's electric or fuel-based should be cleaned and tested monthly. Labtron provides detailed maintenance instructions, troubleshooting tips, and access to service support when needed. With proper care, Labtron portable autoclaves can provide years of reliable service in demanding environments.
Post your question?
Our experts are here to help!
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608