Request A Quote
Portable Electrolyte Analyzer LPEA-A12

Portable Electrolyte Analyzer LPEA-A12

Portable Electrolyte Analyzer LPEA-A12 has the capacity to store up to 5,000 sample records, with options for further data expansion. It functions optimally at temperatures between 5°C and 40°C and relative humidity levels of up to 80%. The analyzer also offers a quality control chart that can be printed directly. It includes special operation instructions and electrode maintenance tips to empower users in troubleshooting. Our electrolyte analyzer, with its user-friendly interface, delivers precise and efficient electrolyte testing in diverse clinical settings.

Request Quote
Sample Type Plasma, Serum, Whole blood, Urine, CSF
Test Parameters K⁺, Na⁺, Cl⁻, Ca, pH
Sample Volume 90µL
Test Range K⁺: 0.50 to 15.00 mmol/L
Na⁺: 20.0 to 200.0 mmol/L
Cl⁻: 20.0 to 200.0 mmol/L
Ca²⁺: 0.3 to 5.0 mmol/L
pH: 6.0 to 9.0
Test Precision K⁺: ≤ 1.0%
Na⁺: ≤ 1.0%
Cl⁻: ≤ 1.0%
Ca²⁺: ≤ 1.5%
pH: ≤ 1.0%
Relative Humidity ≤ 80%
Ambient Temperature 5℃ to 40℃
Atmospheric Pressure 86kPa to 106kPa
Analysis Time 25sec
Power Supply 220V AC, 50Hz
Dimension (L × W × H) 295 × 163 × 293 mm
Weight 3.5 kg
  • Integrated reagent pack with encryption chip
  • 7-inch true-color HD touch screen
  • Minimum sample volume
  • RS232 interface for connection
  • LIS networking
  • USB software upgradeable
  • Supports wired, Bluetooth, and WiFi wireless printers

Our Portable Electrolyte Analyzer LPEA-A12 is used in clinical laboratories and healthcare settings to quickly and accurately measure electrolyte levels (K⁺, Na⁺, Cl⁻, Ca²⁺, pH) in plasma, serum, whole blood, urine, and CSF samples.

K/Na Electrode set for 2 parameters1
K/Na/Cl Electrode set for 3 parameters1
K/Na/Cl/Ca/PH Electrode set for 5 parameters1
Qualification Card1
Cable thermopaper printer1
Standard Accessories Kit 1
Standard Readent Kit 1

 

Standard Accessories Kit:

AccessoryQuantity
Auxiliary pump tube1
Needle for blockage elimination1
AC/DC Adapter1
American Standard Power cord1
European Standard Power cord
British Standard Power cord
India Standard Power cord
Probe tie-in1
Thermopaper1
Electrode O ring5

 

Standard Reagent Kit:

ItemQuantity
Calibration standard Solution (A:650ml; B:100ml)1
Refill Solution for ISE Electrodes (0.8ml*5pcs/set)1
Refill Solution for Ref Electrode (20ml/bottle)1
Multi-enzyme De-proteinizer (2ml*3)1
Conditioner (0.8ml*5pcs/set)1
Electrolytes Quality Control (3High/4Middle/3Low) (1ml*10pcs/set)1
DescriptionPcs/set
Calibration standard Solution (A:650ml; B:100ml)1
Calibration standard Solution (A:850ml; B:200ml)1
K Electrode1
Na Electrode1
Cl Electrode1
Ca Electrode1
pH Electrode1
Reference Electrode1
Electrode Conditioner (5 pieces/set)5
Multi-enzyme De-proteinizer (2ml* 3 pcs/set)3
Refill Solution for ISE Electrodes (5 pieces/set)5
Refill Solution for Ref Electrode (20ml/bottle)1
Probe tie-in1
Auxiliary pump tube1
Electrolytes Quality Control (High/Middle/Low)1
Print Paper Roll1

Models of Portable Electrolyte Analyzer

Labtron Portable Electrolyte Analyzer have compact design for rapid and accurate measurement of electrolyte levels in various samples. These analyzers feature advanced ion-selective electrodes that accurately measure key electrolytes. Some models offer intuitive touchscreen interfaces for easy navigation and data management. With a fast response time, results can be obtained within minutes, making them ideal in clinical settings. They often come with built-in quality control features to ensure reliable results. Our electrolyte analyzer's lightweight and battery-operated design allows for portability, enabling testing in diverse environments.

Portable Electrolyte Analyzer LPEA-A10

Portable Electrolyte Analyzer LPEA-A10

  • Sample Type : Plasma, Serum, Whole blood, Urine, CSF
  • Test Parameters : K⁺, Na⁺
  • Sample Volume : 90µL
  • Test Range : K⁺: 0.50 to 15.00 mmol/L Na⁺: 20.0 to 200.0 mmol/L
Portable Electrolyte Analyzer LPEA-A11

Portable Electrolyte Analyzer LPEA-A11

  • Sample Type : Plasma, Serum, Whole blood, Urine, CSF
  • Test Parameters : K⁺, Na⁺, Cl⁻
  • Sample Volume : 90µL
  • Test Range : K⁺: 0.50 to 15.00 mmol/L Na⁺: 20.0 to 200.0 mmol/L Cl⁻: 20.0 to 200.0 mmol/L

Need a Product

send details for best price quotation

Click
For

Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom

Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608