| Capacity | 270 L |
| Temperature Range | RT +5℃ to +50℃ |
| Saturated Barrel Temperature | +47℃ to +63℃ |
| Temperature Uniformity | ≤±2.0℃ |
| Temperature Fluctuation | ≤±5.0℃ |
| Spray Quantity | 1.0 to 2.0 ml/80cm²/h |
| Power Supply | AC 220 V 50 Hz |
| Internal Dimensions (W × H × D) | 900 × 500 × 600 mm |
| External Dimensions (W × H × D) | 1410 × 910 × 1220 mm |
User-friendly interface
Efficient fog collection
Automated water refill
High-quality nozzle system
Sturdy & long-lasting build
Salt Spray Test Chamber LSSC-C11 is corrosion testing equipment that simulates a saline environment to evaluate the durability of materials and coatings. It is essential for testing the resistance of metals and protective finishes in the automotive, aerospace, and marine industries.
Download catalog for
Salt Spray Test Chamber LSSC-C11 Catalog
Labtron Salt Spray Test Chamber has an adjustable atomizer tower and a spray nozzle spraying system, allowing precise control over spray coverage. These continuous/periodic sprays cover the entire working space, ensuring a diffuse state for even distribution. Their protection against humidifier dry-combustion, over-temperature, over-current, water shortage, and earth leakage ensures safety and reliable operation. They are highly efficient, corrosion-resistant, and designed for a long service life. Our salt spray test chamber has a nichrome heater as the heating element and a saturated air barrel for effective and consistent heating performance.








Labtron Salt Spray Test Chambers are designed to simulate corrosive environments and evaluate the resistance of materials, coatings, and surface treatments against salt-induced corrosion, enabling industries to ensure product durability and compliance with quality standards.
Temperature plays a crucial role in salt spray testing, affecting the evaporation rate, condensation, and overall test consistency. Labtron chambers maintain a stable temperature, typically 35°C (for NSS) and 50°C (for CASS tests), ensuring precise and repeatable results.
Yes, while the standard salt solution is 5% NaCl, different concentrations can be used based on test requirements. Labtron chambers allow adjustable salt concentrations for customized testing according to specific industry standards.
Labtron chambers ensure accurate testing by maintaining the pH level between 6.5 and 7.2 for standard NSS tests. Regular monitoring and adjustments using diluted acetic acid or sodium hydroxide help maintain precise conditions.
Labtron Salt Spray Test Chambers include a corrosion-resistant testing chamber, salt solution reservoir, air saturator, spray nozzles, temperature controller, and a sealed enclosure, all working together to maintain a controlled test environment.
Test specimens should be placed at a 15° to 30° angle to ensure even exposure to the salt mist. Labtron chambers include customizable racks and holders for proper sample positioning based on test standards.
Yes, Labtron chambers can be used alongside tests like humidity testing, cyclic corrosion testing, and outdoor exposure tests for a more comprehensive corrosion resistance evaluation.
Labtron chambers typically operate on single-phase or three-phase power, depending on the chamber size and model. Specific voltage and power requirements are provided in the product manual.
Post your question?
Our experts are here to help!
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608