Capacity | 32 L |
Temperature Range | 0℃ to 80℃ |
Blackboard Temperature | +45℃ to +90℃ (±3℃) |
Temperature Deviation | ±2℃ (without lighting) |
Temperature Resolution | 0.1℃ |
Temperature Fluctuation | ±0.5℃ (without lighting) |
Effective Irradiation Area | 625 cm² |
Door Opening Method | Single door, opens from right to left |
Sample Holder | Flat, height-adjustable, suitable for special-shaped samples |
Sample Space | Three-quarters of the total space; no heat load |
Humidity Range | 20% to 95% R.H |
Humidity Deviation | ±3% R.H (without lighting) |
Humidity Fluctuation | ±2.5% R.H (without lighting) |
Light Source | Full-spectrum air-cooled Xenon lamp |
Lamp Power | 1.8 KW |
Sensor | Irradiance sensor |
Irradiance Monitoring Method Narrow Band | 340 nm / 420 nm |
Irradiance Monitoring Method Broad Band | 280 to 800 nm / 300 to 400 nm |
Adjustable Light Irradiance Range | 0.35 to 0.70 W/m² (at 340 nm) 0.35 to 1.25 W/m² (at 420 nm) 180 to 1200 W/m² (at 280 to 800 nm) 30 to 60 W/m² (at 300 to 400 nm) |
Spray Function | Indirect continuous spray; manual pure water supply |
Function Modes | Alternating: Irradiation mode, Dark mode, Spray mode |
Ambient temperature | 25℃ ±3℃ |
Condenser | 5 to 25℃ ±3℃ |
Refrigerant | R404a |
Relative humidity | ≤85% |
Air pressure | 86 kPa to 106 kPa |
Inner Material | SUS#304 mirror stainless steel |
External Material | Cold-rolled steel plate with spray paint |
Duct Circulation | Centrifugal circulation motor |
Door Lock | Flat handle lock; door opens from left to right |
Controller | 7-inch touch screen controller Power switch Running indicator USB data download interface |
Power Cable | Equipped with drain outlets and a 2.5 m national standard cable at the back |
Water Supply | Pure water or distilled water manually added into the water tank Customization for auto water supply available |
Frequency | 50 Hz (Customization for 60 Hz available) |
Protection Grounding Resistance | ≤4Ω |
Maximum Current | 10A |
Power Consumption | 4.0 KW |
Power Supply | AC 380 V ±10% with protection grounding (Customization for 220 V available) |
Interior Dimension | 350 × 300 × 300 mm |
Exterior Dimension | 900 × 900 × 770 mm |
Packaging Dimension | 900 ×900 × 102 mm |
Gross Weight | 196 Kg |
Net Weight | 142 Kg |