| Tested Opening | 180 mm |
| Maximum Opening | 530 mm |
| Airflow System | 70% air recirculation, 30% air exhaust |
| Average Inflow Velocity | 0.50 m/s |
| Average Downflow Velocity | 0.33 m/s |
| Inflow Airflow Volume | 549 m3/h |
| Downflow Airflow Volume | 1277 m3/h |
| Exhaust Airflow Volume | 549 m3/h |
| Filter Specification | ULPA > 99.999% efficiency at 0.3 µm |
| Cleanliness Grade | Class 3 (ISO 14644.1) |
| Display | 7 inches touch screen |
| Sash Glass Type | Tempered Glass, UV-proof |
| Fluorescent Lamp | 30W |
| UV Lamp | 3OW |
| Fluorescent Light Intensity | ≥700 lux |
| Sound Emission | ≤65 Db |
| Colony Count (Cross Contamination Protection) | ≤2 CFU/Time |
| Material | External: 18-gauge Electro-galvanized steel with white oven-baked powder-coated finish; Work Surface: 16-gauge Stainless steel |
| Power Consumption | 1500W |
| Power Supply | AC 220, 50/60HZ |
| Internal Dimension (W × D × H) | 1600 × 672 × 630 mm |
| External Dimension (W × D × H) | 1780 × 790 × 2080 mm |
| Packaging Dimension (W × D × H) | 1910 × 950 × 2220 mm |
| Net Weight | 300 Kg |
| Gross Weight | 340 Kg |
Anti-Vibration Structure
Filter Replacement Indicator
Premium Stainless-Steel
Recirculation and Exhaust
Accurate Airflow Control
Our biosafety cabinet is suitable for sterile pharmaceutical handling, environmental testing, and quality control in the food and agricultural biotech industries, ensuring a contaminant-free environment for sensitive research, diagnostic applications.
Fluorescent Lamp
Ultraviolet Lamp
Receptacle
Detachable Base Stand
Accessories | Quantity |
Gas Valve | 1 |
Vacuum Valve | 1 |
Water Valve | 1 |
Arm Rest | 1 |
I.V. Pole | 1 |
Canopy Connection | 1 |
Download catalog for
Biosafety Cabinet Class II A2 LBS2-A52 Catalog
Available Range :
Biosafety Cabinet Class II A2 LBS2-A50 : 180 mm Biosafety Cabinet Class II A2 LBS2-A41 : Safety Height 200 mm Biosafety Cabinet Class II B2 LBS2-B21 : Safety Height 200 mm(8 inch) Biosafety Cabinet Class II B2 LBS2-B20 : Safety Height 200mm (8 inch) Biosafety Cabinet Class II B2 LBS2-B24 : Safety height 200mm (8'') Biosafety Cabinet Class II A2 LBS2-A20 : Safety height of 200 mm (8 inch)
Labtron Biosafety Cabinets are incorporated with advanced technology, including ULPA filters and UV indicators. These cabinets ensure advanced protection with built-in interlock functionalities. Our diverse selection encompasses Class I, Class II A2, Class II B2, and more. They provide exceptional airflow control for maintaining sterility and optimizing laboratory conditions. Our biosafety cabinet provides unparalleled durability, featuring two-layer laminated toughened glass.








The digital control panel on the Biosafety Cabinet Class II A2 LBS2-A52 provides real-time monitoring of key parameters such as airflow velocity, filter life, and UV lamp status. This allows users to easily ensure the cabinet is functioning optimally without guesswork. The touchscreen interface is user-friendly and enables quick adjustments to settings. Password protection prevents unauthorized changes, maintaining operational security. Audible and visual alarms integrated into the system alert users immediately to any malfunctions or safety concerns. Overall, the control panel streamlines operation while enhancing safety and convenience.
Regular maintenance of the Biosafety Cabinet Class II A2 LBS2-A52 includes periodic filter inspections and timely replacements to maintain ULPA filtration efficiency. The removable stainless steel work surface should be cleaned and sterilized routinely to prevent contamination buildup. The motorized sash and glass panels require cleaning with appropriate non-abrasive agents to preserve visibility and function. Users should routinely test the UV lamp and replace it as needed to ensure effective sterilization. Electrical components should be checked for moisture damage due to the cabinet’s waterproof design. Adhering to these maintenance steps ensures consistent performance and extends the cabinet’s service life.
The Biosafety Cabinet Class II A2 LBS2-A52 is engineered to operate quietly, with noise levels maintained at or below 65 decibels. This low noise output helps create a comfortable working environment, especially during long experiments. The cabinet’s high-performance blower is designed for noise reduction without sacrificing airflow stability. Quiet operation reduces distractions and helps maintain concentration for sensitive tasks. The cabinet’s sound insulation also minimizes disturbance to adjacent lab areas. Users benefit from an optimal balance of performance and comfort.
Yes, the Biosafety Cabinet Class II A2 LBS2-A52 offers options for integration with external monitoring and laboratory information management systems (LIMS). This integration allows remote monitoring of cabinet performance, including airflow, filter status, and alarm conditions. Data logging capabilities support compliance with quality assurance and audit requirements. Integration helps streamline lab workflows and enhances traceability of biosafety cabinet usage. It also enables centralized control of multiple cabinets in large facilities. Such connectivity ensures proactive maintenance and improved safety management.
The integrated UV lamp features a programmable timer that allows precise control over sterilization cycles. This prevents overexposure, which can degrade cabinet materials and pose health risks. Timed sterilization ensures thorough decontamination of the work surface between uses. Users can schedule UV activation during off-hours for convenient and safe cleaning. The UV system complements airflow filtration to provide a dual layer of contamination control. Proper use of the timer extends the lamp’s lifespan while maintaining effective sterilization.
Post your question?
Our experts are here to help!
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608