| Wavelength Range | 190 to 900nm |
| Spectral Bandwidth | 0.1, 0.2, 0.4, 1.0, 2.0nm (automatic adjustable) |
| Wavelength Accuracy | ±0.15nm |
| Wavelength Repeatability | ±0.1nm |
| Baseline Stability | ≤0.002A/30 minutes (static); ≤0.005A/30 minutes (dynamic) |
| Grating | 1800mm |
| Environment Temperature | +15°C to +35°C |
| Relative Humidity | 20% to 80% |
| Light Source | 6-lamp automatic turret, automatic alignment |
| Characteristic Concentration (Cu) | 0.015μg/mL/1% |
| Detection limits (Cu) | 0.002μg/mL |
| Precision | RSD≤0.5% |
| Background Correction | Correction of 1A background |
| Measurement Methods | Flame method, hydride method |
| Repetition Frequency | 1 to 99 times |
| Interface | Computer and USB interface communication |
| Power Consumption | 6kW |
| Power Supply | 220V (±5% to -10%), 50/60 Hz; 5000VA |
| Product Dimension (L×W×H) | 990 mm× 430 mm× 410 mm |
| Packaging Dimension (L×W×H) | 1260 mm× 640 mm× 70 mm |
| Net Weight | 100kg |
| Gross Weight | 120kg |
Efficient glass atomizer with spray chamber
Includes Metal titanium combustion head
Double cathode power built-in high-performance lamps
Automatic PC control alignment, optimization, ignition
Automatic safety protection function
Anti-temperature/automatic gas path protection
Deuterium background correction
Our Atomic Absorption Spectrophotometer LAAS-A26 precisely measures trace metal concentrations in pharmaceuticals, food safety, environmental monitoring, and industrial quality control, ensuring accurate, reliable, and reproducible analytical results.
HCL Lamp (Pb,Cd,Cu) 3pc
Hollow cathode lamp
Graphite tube
Main unit dust cover
3A fuse wire
1.5A fuse wire
Three core power line
Main Unit Signal Line
Plastic Pipe
Screwdriver
Acetylene hose clip
Acetylene gas pipe
Desktop Computer
Printer
Oil-less Air Compressor
Graphite System
HCL Lamp (Se, Fe, Mn, Zn, Na, K)
Hydride Generator
Water Chiller
Cooling Water System
Auto Sampler (Flame)
Auto Sampler (Graphite Furnace)
Auto Sampler (Flame and Graphite Furnace)
Glass Atomizer
Adjustable Atomizer (1-10 mL)
Hydrofluoric Acid Resistant Type Atomizer
Electrically Heated Quartz Tube (Normal)
Electrically Heated Quartz Tube (White Porcelain)
Download catalog for
Atomic Absorption Spectrophotometer LAAS-A26 Catalog
Available Range :
Atomic Absorption Spectrophotometer LAAS-A21 : 180 – 900 nm Atomic Absorption Spectrophotometer LAAS-A23 : 185 to 900 nm Atomic Absorption Spectrophotometer LAAS-A20 : 190 to 900 nm Flame Atomic Absorption Spectrophotometer LAAS-A15 : 190 to 900nm Atomic Absorption Spectrophotometer LAAS-A24 : 190nm to 900nm Atomic Absorption Spectrophotometer LAAS-A30 : 3600 L/mm: 190 nm - 500 nmLabtron Atomic Absorption Spectrophotometers offer precise and efficient trace element analysis with adaptable atomization technologies and high-precision optical systems. These feature a titanium combustion head, polymer explosion-proof spray chamber with enhanced safety features and durability. We offer versatile operation modes, including flame, graphite furnace, and hydride generation systems, these meet diverse analytical needs. Our Spectrophotometers boast powerful software and user-friendly design for consistent, reliable results and seamless data management.








An analytical tool called an Atomic Absorption Spectrophotometer (AAS) uses the light that each element's atom absorbs to determine the concentration of that element in a sample.
The basis of spectrophotometry is the light uptake by unbound atoms in a gaseous state. An excited state is created when atoms in the ground state absorb light energy at particular wavelengths. Analyte content in the sample can be calculated by determining the amount of light consumed.
The components include an atomizer (commonly used atomizers are Flame atomizers and Electro thermal atomizers), a Radiation source, and a spectrophotometer.
The hollow cathode lamp in atomic absorption spectrophotometer emits distinctive light at particular wavelengths that match the target analyte. This guarantees that the analyte atoms are excited selectively, resulting in exact and reliable readings.
Labtron’s Atomic Absorption Spectrophotometers boast flexible atomization technologies, accurate optical systems, a titanium combustion head, and a polymer explosion-proof spray chamber. These elements guarantee accurate and effective trace element analysis, along with improved safety and longevity.