| Measuring Range | 40 to 370℃ (104 to 698℉) |
| Resolution | 0.1℃ |
| Heater Rate | Procedure A: 5℃/min to 6℃/min Procedure B: 1.0℃/min to 1.5℃/min Procedure C: 2.5℃/min to 3.5℃/min |
| Stirring Speed | Procedure A: 90r/min to 120r/min Procedure B: 250r/min ± 10r/min Procedure C: 90r/min to 120r/min |
| Sample Preheating | Side: Ambient to 95℃; Center: Ambient to 95℃ |
| Safety Features | Automatically shuts down, alarm, and cooling, in case: Flash Point Mode: EFP + 20℃ is reached 400℃ is reached |
| Cooling | Air cooling |
| Gas Supply | LP gas, butane gas, pipeline gas or natural gas (pressure < 10kPa) |
| Ignition Source | Gas ignition or electric ignition, interchangeable |
| Ignition Power | ≤50W |
| Flash Point Detection | Thermocouple |
| Temperature Sensor | PT100 probe |
| User interface | 7 |
| Languages | Chinese, English, Russian |
| Barometric correction | Flash point is corrected to barometric pressure |
| Data Storage | No less than 5000 results |
| I/O Port | RS232, RJ-45 |
| Standards | ASTM D93, ISO 2719, IP 34, GB/T 261 |
| Data Export | LIMS or Printer |
| Machine Power | 800W |
| Heater Power | 500W |
| Operating Conditions | 5℃ to 35℃ 20 to 90%RH at 35℃ Non-condensing No air flow |
| Power Supply | AC 220V±10 50/60Hz |
| Dimensions (W×D×H) | 265mm× 525mm× 365mm |
| Weight | 20 Kg |
| Gross Weight | 35 Kg |
Our PMCC Flash Point Tester LCFP-A15 is essential in laboratories, oil refineries and chemical manufacturing. It ensures compliance with safety regulations and effective risk management in hazardous materials handling.
Download catalog for
PMCC Flash Point Tester LCFP-A15 Catalog
Labtron Closed Cup Flash Point Tester is specialized to determine lowest temperature at which vapor above liquid can ignite when exposed to open flame or spark. They are equipped with thermometers to ensure readings upto 300℃ during testing. Their sample containers are made from heat-resistant materials to withstand high thermal condition. They include built-in ventilation system that manage fumes generated during process. Our Closed Cup Flash Point Tester to assess flammability risk of fluids in petrochemicals, paints, and solvents.

.webp)




Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608