Microplate Reader LMPR-A10 has a wavelength range of 400 to 750 nm for accurate processing of biological samples. It is designed with a 96-well plate for high-throughput analysis. Our unit employs a bichromatic multi-channel optical system to measure samples for enhanced sensitivity. It includes a linear shaking function that allows precise sample control with 3 variable speeds. This unit offers qualitative and quantitative data evaluation for reliable results.

Request Quote
Plate Type 96-well plate
Wavelength Range 400 to 750 nm
Standard Wavelength 405, 450, 492, 630 nm
Wavelength Accuracy <2 nm
Wavelength Precision ± 0.03 A
Resolution 0.001 Abs
Wavelength Linearity R ≥ 0.995
Wavelength Repeatability < 0.5% at 2.0 OD
Wavelength Stability < 0.005 Abs/10min
Reading Speed 5s single wavelength
Reading Range 0.000 to 4.000 Abs
Measurement System 8-Channel Optical System
Measurement Range 0.001 to 2.500 Abs
Measurement Speed 6 to 10 sec for 96 well plate
Photometric Range 0.000 to 3.500Abs
Photometric Accuracy < 1% or ± 0.001Abs
Photometric Repeatability < 1%
Temp. Range 10°C to 35°C
Humidity Range 15% to 85%
Shaking Linear (slow, medium and high mode)
Light Source Monochromatic LED
Light Source Life More than 100000 h
Detection Method Absorbance
Detector Silicon Photodiode
Detection Modes Comprehensive Qualitative and Quantitative Data-Evaluation with Cut-Offs, Curve-Fits and Transformation Formulas
Programs 55 user programs
Power Consumption 80 W
Power Supply 240 V, 60 Hz
Dimension 400 × 310 × 169mm
Net Weight 6 kg
Gross Weight 10 kg
  • Rapid analysis
  • 7-inch TFT LCD Display
  • Can store up to 1200 results
  • Malfunction alarm system
  • Two free positions for additional filters

Microplate Reader measures the optical properties of samples. It is ideal for life science research, pharmaceutical laboratories, and clinical diagnostics for accurate data analysis and results.

Similar Products


Microplate Reader LMPR-A10 FAQ's
  • Microplate Reader LMPR-A10 comes with an integrated, user-friendly software with 55 programs to help you ease the process of data analysis. The data can be exported in .csv or .exe file formats.

  • Our Microplate Reader LMPR-A10 can be used for studies having one particular application which demands high repeatability and sensitivity. It is suitable for monitoring of microbial growth which takes more than one day, and thioflavin T experiments which takes up to 7 days.

  • A Microplate Reader, commonly called as plate reader or plate spectrophotometer, is an analytical instrument that reads or detects the physical, chemical, and biological events of the analyte, within the wells of the plate. It has numerous tiny wells where the sample analyte is introduced for analysis.

  • A standard spectrophotometer measures the absorbance of one sample at a time. The sample is typically placed in a cuvette through which light is sent horizontally. An absorbance plate reader offers higher throughput and can measure the absorbance of samples in microplates (typically 96-well or even 384-well) by sending light through each well vertically.

  • Microplate-based measurements detect light signals produced by a sample, converted by a sample, or transmitted through a sample. In the plate reader, the signal is measured by a detector, usually a photomultiplier tube (PMT). PMTs convert photons into electricity that is then quantified by the microplate reader. The output of this process is numbers by which a sample is quantified.

  • Tungsten-halogen or xenon flash lamps are generally used. LEDs have also come into the picture in some modern readers.

  • The detection technology can be of 3 types- Monochromator based detection which provides flexibility, convenience, and spectral scanning; Filter based detection which enables rapid switch between different wavelengths; Hybrid detection which has dual technological advantage of both flexibility and sensitivity.

Post your question?

Our experts are here to help!

Microplate Reader : Labtron Microplate Readers have multiple wells (commonly 48, 96, 384, or 1536) that serve as small test tubes. These are used to detect biological, chemical, or physical events in samples within microtiter plates. Our advanced models offer a range of wavelengths, allowing users to customize the absorbance or fluorescence measurements according to specific requirements. With their high sensitivity and dynamic range, they are capable of detecting very low and high signals. They come with specialized software for controlling the reader, setting up protocols, and analyzing the collected data. To ensure that reagents and samples are well-mixed before measurement, our microplate readers include shaking or stirring functions.
Microplate Reader LMPR-A12, Microplate Reader LMPR-A13, Microplate Reader LMPR-A14, Microplate Reader LMPR-A15, Microplate Reader LMPR-B10, Multi-Mode Microplate Reader LMMR-A10, Microplate Reader LMPR-A30, Microplate Reader LMPR-A20, and more.

Quick Links : About Us, Contact Us, User Manuals, Career, Terms of Use, Policies, Sitemap, Gallery.

Need a Product

send details for best price quotation

Click
For

Address: Labtron Equipment Ltd. Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom

Email: info@labtron.com Whatsapp: +44 20 8004 3587 Phone: +44 2080 043608

Copyright © 2024 Labtron Equipment Ltd. All rights reserved.