Request A Quote
Automatic Cleveland Open Cup Flash Point Tester LOFP-A12

Automatic Cleveland Open Cup Flash Point Tester LOFP-A12

Automatic Cleveland Open Cup Flash Point Tester LOFP-A12 operates efficiently within an ambient temperature range of 10°C to 40°C and a relative humidity of up to 80%. It allows presetting of flash point temperature and other parameters. The built-in printing function enables quick data output. This device automatic temperature control system enhances accuracy. Our flash point tester delivers precise and reliable flash point measurements for petroleum-based substances. 

Request Quote
Ambient Temperature 10 ℃ to 40 °C
Relative Humidity ≤ 80 %
Temperature Measurement- Full Scale Ambient temperature to 300 °C
Temperature Measurement- Repeatability ≤ 8 °C
Temperature Measurement- Reproducibility ≤ 16 °C
Temperature Measurement- Resolution 0.1 °C
Test accuracy 0.002
Precision 0.005
Flame 3 to 4 mm
Power Consumption ≤ 500 W
Power Supply AC 220 V ± 10 %, 50 Hz
Dimensions 550 x 470 x 410 mm
  • Clear LCD display

  • User-friendly interface

  • Compact and durable design

  • Automatic test data printing

  • Electric flame ignition

Our Automatic Cleveland Open Cup Flash Point Tester LOFP-A12 is used to determine the flash point of petroleum products and bituminous materials. This tester is ideal for laboratories, refineries, and research facilities. 

Models of Open Cup Flash Point Tester

Labtron Open Cup Flash Point Tester features adjustable heating power for precise control over temperature settings, with an electric current meter to display power consumption. These include menu prompts and guided input functions, simplifying the setup process for users. Their electric ignition mode ensures a reliable start-up for smooth operation. They have a detachable ionization ring for easy cleaning, maintenance, and longevity. Our open cup flash point tester has a barometric correction for temperature calibration, ensuring accurate readings and accounting for atmospheric pressure variations.

Open Cup Flash Point Tester LOFP-A10

Open Cup Flash Point Tester LOFP-A10

  • Thermometer : 0 ℃ ~ 360 °C
  • Scale Division : 1 °C
  • Ambient Temperature : ≤ 35 °C
  • Relative Humidity : ≤ 85 %
Automatic COC Flash & Fire Point Tester LOFP-B10

Automatic COC Flash & Fire Point Tester LOFP-B10

  • Ambient Temperature : 15 ℃ ~ 35 ℃
  • Relative Humidity : ≤ 85 %
  • Temperature Measurement- Full Scale : Ambient temperature to ~ 400 ℃
  • Temperature Measurement- Repeatability : ≤ 4 ℃
Automatic COC Flash Point & Fire Point Tester LOFP-B11

Automatic COC Flash Point & Fire Point Tester LOFP-B11

  • Measuring Range : 79 to 370℃ (175 to 698℉)
  • Resolution : 0.1℃
  • Repeatability : ≤ 6℃
  • Reproducibility : ≤ 8℃
Cleveland Open Cup Flash Point Tester LOFP-A11
Cleveland Open Cup Flash Point Tester LOFP-A11
  • Temperature Measurement Range : Room Temperature to 400℃
  • Repeatability : Flash Point ≤150℃: ±2℃ Flash Point ≥150℃: ±3℃
  • Resolution : 0.1℃
  • Precision : 0.5%
Automatic Cleveland Open Cup Flash Point Tester LOFP-A12 FAQ's
  • The Automatic Cleveland Open Cup Flash Point Tester LOFP-A12 uses a small open flame for ignition, which is compatible with liquefied petroleum gas (LPG), butane, natural gas, or pipeline gas. The gas should be clean, dry, and regulated at a pressure below 10?kPa to ensure proper flame generation. Using contaminated or unregulated gas can result in ignition failure or unstable flame size. We recommend performing gas line checks and leak tests before each use.

  • Our Automatic Cleveland Open Cup Flash Point Tester LOFP-A12 may produce unreliable or abnormally low flash point readings if the sample contains water, dust, or volatile impurities. These contaminants can lead to premature ignition, posing safety hazards and resulting in invalid data. We recommend that all samples be clean, homogenous, and prepared according to standards like ASTM or ISO. Contaminated samples can also cause buildup or damage to the test cup and flame sensor.

  • Labtron advises that the Automatic Cleveland Open Cup Flash Point Tester LOFP-A12 be operated only in well-ventilated areas, far from any external ignition sources or flammable materials. Users should wear gloves and safety goggles when handling petroleum-based samples. Before testing, check all gas line connections for leaks and ensure the flame ignition area is clear. The tester should never be left unattended during a test, and users must follow emergency shutdown procedures in case of flame failure or system fault.

  • The Automatic Cleveland Open Cup Flash Point Tester LOFP-A12 commonly encounters user errors such as overfilling the sample cup, improper sample mixing, incorrect flame size, and failure to input or verify atmospheric pressure. These mistakes can lead to inaccurate or inconsistent flash point results, compromising test reliability. We advise users to follow the standard operating procedure (SOP) closely, conduct thorough pre-test equipment checks, and pay attention to system prompts and alerts. Regular training sessions and periodic refresher programs are also recommended to minimize operator-related errors.

  • Labtron Automatic Cleveland Open Cup Flash Point Tester LOFP-A12 may display an error code or flame failure warning if there is an issue during testing. In such cases, users should first pause the test and allow the unit to cool down safely. We recommend checking the ignition system, confirming a stable gas supply, and removing any blockages or residue from the flame nozzle area. Refer to the error code guide in the user manual to troubleshoot accurately. If the issue continues, the test should be restarted with a fresh sample, and users should contact our technical support for further assistance.

Post your question?

Our experts are here to help!

Need a Product

send details for best price quotation

Click
For

Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom

Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608

Copyright © 2026 Labtron Equipment Ltd. All rights reserved.