| Drying Area | 0.4 m² |
| Shelf Temperature | -50 to +60 ℃ |
| Shelf Temperature Limit | Less than or Equal to -55℃ (no-load) |
| Shelf Temperature Uniformity | ±1℃ (under stabilized state) |
| Shelf specification | 335 × 300 mm |
| Number Of Shelves | 4 Layers |
| Shelf Spacing | 65 mm |
| Vial Loading Capacity Ø16 | 1650 |
| Vial Loading Capacity Ø22 | 870 |
| Bulk Drug Loading Capacity | 4 kg/batch |
| No-load Temperature | Min.to Less than or Equal - 86℃ (cold trap) |
| Ice Holding Capacity | 10 Kg |
| 24-hour Maximum Ice Holding Capacity | 3 Kg |
| Cold Trap Volume | 21 Liters |
| No-load Vacuum Pressure | 5 Pa |
| Ultimate Vacuum | 1Pa |
| Cooling Method | Air Cooling |
| Ambient Temperature | Less than or Equal to 26℃ |
| Installed Capacity | 6.5 Kw (380V 50Hz) |
| Dimension | 800 × 800 × 1550 mm |
| Weight | 300 Kg |
Pilot Top Press Freeze Dryer LPFD-A14 is used in pharmaceutical and biotech industries for freeze-drying injectable drugs, vaccines, and heat-sensitive biological materials.
Download catalog for
Pilot Top Press Freeze Dryer LPFD-A14 Catalog
Available Range :
Pilot Standard Freeze Dryer LPFD-G10 : 0.16 ㎡ Pilot Standard Freeze Dryer LPFD-F20 : 0.2 m² Pilot Standard Freeze Dryer LPFD-G11 : 0.24 ㎡ Pilot Standard Freeze Dryer LPFD-F21 : 0.3 m² Pilot Standard Freeze Dryer LPFD-F22 : 0.4 m² Pilot Standard Freeze Dryer LPFD-B12 : 0.54 m² Pilot Standard Freeze Dryer LPFD-B13 : 0.82 m² Pilot Standard Freeze Dryer LPFD-B14 : 1 m² Pilot Standard Freeze Dryer LPFD-B15 : 1.15 m²Labtron Pilot Freeze Dryer is specialized to remove moisture from material while preserving its structure and biochemical properties. They are equipped with condenser working at temperature ranging from -70℃ to -80℃. Their automatic defrost cycles reduce manual intervention and streamline maintenance procedures. They utilize vacuum technology to preserve quality of sensitive materials during drying process. Our Pilot Freeze Dryer is efficient for pharmaceuticals, food storage, and research applications.








Labtron’s Pilot Top Press Freeze Dryer is specifically designed for use in the pharmaceutical and biotech industries. It is well-suited for freeze-drying injectable drugs, vaccines, and sensitive biological materials. The system ensures product safety and high-quality drying by offering precise temperature control through silicone-oil-based circulating shelves and close vacuum monitoring. Its ability to store multiple recipes and handle bulk drug or vial-based formats makes it ideal for small-scale production and process optimization before scaling up.
Temperature control is critical for the freeze-drying of sensitive materials, and Labtron addresses this with circulating heat transfer fluid inside the shelves. This design allows both heating and cooling to be applied uniformly across all layers, helping maintain consistent product temperature during each stage. The shelf temperature remains stable within ±1°C under steady-state conditions, which ensures better drying results and prevents damage to delicate samples.
Safety is built into multiple parts of the Labtron freeze dryer. It features real-time vacuum pressure monitoring to detect system issues early. If the vacuum level doesn’t reach its setpoint in time, the system automatically shuts down the vacuum pump and opens a solenoid valve to equalize pressure, protecting both the machine and the samples. Additionally, a molded T-structure sealing system prevents leaks, and the compressor includes a restart delay and thermal overload protection to avoid overheating.
Yes, the Labtron freeze dryer offers high customization flexibility. Users can store up to 50 recipes, each with 36 temperature stages, and assign unique names to them. Time and temperature can be modified in real time while the process is running. The intelligent control system waits for temperature setpoints before starting countdowns, and it automatically adjusts the drying time if parameters are changed mid-process. This allows researchers and manufacturers to fine-tune freeze-drying protocols for different products.
The freeze dryer is designed for efficient lab or pilot-scale use. It has a 0.4 m² drying area, 10 kg ice holding capacity, and can handle both vials (up to 1,650 with Ø16mm) and bulk drug loads of 4 kg per batch. It reaches vacuum levels as low as 1 Pa and cold trap temperatures down to -86°C. The compact unit is easy to install in a standard lab, and its air-cooling system ensures quiet and stable operation. Combined with user access control and a 7-inch touchscreen for easy monitoring, it offers a complete package for advanced freeze-drying needs.
Post your question?
Our experts are here to help!
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608