Request A Quote
Analytical HPLC System LAHS-E10

Analytical HPLC System LAHS-E10

Analytical HPLC System LAHS-E10 deliver a flow rate range of 0.0000 to 12.0000 mL/min  ensuring precise solvent delivery for diverse analytical applications. The system offers a comprehensive detector range including UV, ECD, FLD, ELSD, and RID managed with intelligent control. It features obstacle protection and empty bottle detection to prevent needle damage and invalid sample injection. The system uses innovative technology to identify shoulder and small peaks in chromatographic analysis. Our Analytical HPLC is ideal for precise separation and detection of complex mixtures in pharmaceuticals, chemicals, and biotechnology research.

Request Quote
Flow Rate Setting Range 0.0000 to 12.0000mL/min
Flow Rate Setting Increment 0.0001mL/min
Gradient Error Less than equal to 0.75%
Gradient Stability 0.2% RSD (5% to 95%, 0.2 to 2.0mL/min)
Flow Rate Setting Value Error ± 0.20%
Flow Rate Stability Less than equal to 0.075% RSD
Maximum Operating Pressure 62MPa /75MPa
Autosampler Linearity R ≥ 0.9999 (1.0μL to 50μL)
Autosampler Cross Contamination Less than equal to 0.002%
Number Of Injection Bits 108 bits (2 ×54, 2mLsample bottle)
Injection Mode flow to needle
Injection Volume 100μL, 200μL, 500μL, 1000μL, 1500μL, 2000 μL
Repeatability Less than 0.3 % RSD (2μL to Max)
Less than 1.0 % RSD (0.1μL to 1.9μL)
Linear Correlation Coefficient Greater than 0.9999
Number of Sample Position 108 (2×54, 2mL), 96-well plates
Temperature Setting Range 4 to 40 ℃ (optional refrigeration)
Temperature Setting Value Error ±0.5℃ (optional refrigeration)
Temperature Stability ±0.5℃ (optional refrigeration)
Column Oven Temperature Range 4℃ to 90℃
Column Oven Temperature Stability ± 0.03℃/h
Column Oven Temperature Setting Value Error ± 0.1℃
Power Supply 220V ±10%, 50–60Hz
Autosampler Dimension 930 × 520 × 570 mm
Column Oven Dimension 690 × 520 × 550 mm
Pump & Detector Dimension 830 × 520 × 430 mm
Fluorescence Detector Dimension 640 × 480 × 400 mm
Refractive Index Detector Dimension 640 × 470 × 400 mm
Electrochemical Detector Dimension 610 × 520 × 370 mm
Autosampler Weight 51 Kg
Column Oven Weight 32 Kg
Pump & Detector Weight 32.2 Kg
Fluorescence Detector Weight 37.5Kg
Refractive Index Detector Weight 29.5 Kg
Electrochemical Detector Weight 28.5 Kg
Total Weight 232 Kg
  • Modular system design

  • Multiple detector options

  • Optimized needle cleaning

  • User-friendly workstation

  • Automated injection control

Analytical HPLC System LAHS-E10 is ideal for quantitative and qualitative analysis in pharmaceuticals, biotechnology, and chemical industries. It is widely used for purity testing, component identification, and trace compound detection in complex mixtures.

Accessories 

Quantity 

Unit 

Column Oven

1

Set 

Columns

3

Pcs

Solvent Organizer

1

Piece

Accessories

Quantity

Unit

FLD3400 Detector Flow cell assembly

1

Piece

FLD3400 Detector Xenon lamp

1

Piece

RID3400 Detector Flow cell assembly

1

Piece

ECD3100 Detector

1

Piece

Pump High pressure sealing ring

2

Pieces

Pump Main pump inlet valve assembly

2

Pieces

Pump Auxiliary pump inlet valve assembly

2

Pieces

Autosampler High pressure sealing ring

1

Piece

Autosampler Low pressure sealing ring

1

Piece

Autosampler Injection needle assembly

1

Pack

Accessories Aqueous phase needle filter

1

Pack

Accessories Organic phase needle filter

1

Pack

Accessories Sample bottle, cap

2

Pieces

Accessories Filter sieve plate

1

Piece

Accessories 1/16 PEEK hand tight fittings

8

Pieces

Accessories PEEK pipe

2

meters

Accessories PTFE pipe-ID0.25

2

meters

Accessories PTFE pipe-ID0.75

1

meter

Models of Analytical HPLC

Labtron Analytical HPLC ensure separation of components based on their chemical properties in sample. Their high resolution enhances detection of minute differences in compound concentrations. They employ elevated pressure pump for efficient solvent movement through chromatography column. Their automated sample injection system increases throughput, reducing need for manual handling. Our Analytical HPLC is ideal in pharmaceutical, environmental, and chemical analysis for quality control, research, and regulatory testing.

Analytical HPLC System LAHS-B10

Analytical HPLC System LAHS-B10

  • Wavelength Range : 190 to 700 nm
  • Pump Type : Double plunger reciprocating pump (1 Set)
  • Flow Range : 0.001 to 10.00 mL/min
  • Setting Step Size : 0.001 mL/min
Analytical HPLC System LAHS-B11

Analytical HPLC System LAHS-B11

  • Pump Type : Double plunger reciprocating pump
  • Pump Sets : 2
  • Flow Range : 0.001 to 10.00 ml/min
  • Pump Setting Step Size : 0.001 ml/min
Analytical HPLC System LAHS-D10

Analytical HPLC System LAHS-D10

  • Pump Flow Range : 0.001 to 10.000mL/min
  • Pump Set Step : 0.001mL/min
  • Pump Pressure Range : 0 to 42 MPa
  • Pump Pressure Pulsation : 0.1MPa (10MPa, C18, methanol 1mL/min)
Analytical HPLC System LAHS-B12
Analytical HPLC System LAHS-B12
  • Pump Type : High pressure constant flow pump
  • Solvent Channel : 2
  • Flow Range : 0.001 to 10.000 mL/min
  • Pump Flow Accuracy : Up to ± 0.3 %
Analytical HPLC System LAHS-C10
Analytical HPLC System LAHS-C10
  • Wavelength Range : 190 to 800 nm
  • Pump Flow Range : 0.001 to 10.000 mL/min (step 0.001 mL/min)
  • Gradient Accuracy : Up to ± 1%
  • Retention Time Repeatability : Up to 0.1%
Analytical HPLC System LAHS-A11

Analytical HPLC System LAHS-A11

  • Flow Range : 0.001 to 10.000 mL/min (step size 0.001 mL/min)
  • Flow Rate Accuracy : ≤ ±0.2 %
  • Flow Stability (RSD) : ≤ 0.06 %
  • Maximum Working Pressure : 63 MPa
Analytical HPLC System LAHS-A10

Analytical HPLC System LAHS-A10

  • Pump Type : Quaternary Low-Pressure Pump
  • Pump Flow Rate Range : 0.001 to 10.000 ml/min, in 0.001ml increments
  • Pump Flow Accuracy : ≤ ± 0.2% (1.00 ml/min, H2 O, R.T., 8.5 MPa)
  • Pump Flow Precision : RSD ≤ 0.06% (1.00 ml/min, H2 O, R.T., 8.5Mpa)

Analytical HPLC:
Analytical HPLC System LAHS-B10, Analytical HPLC System LAHS-B11, Analytical HPLC System LAHS-D10, Analytical HPLC System LAHS-B12, Analytical HPLC System LAHS-C10, Analytical HPLC System LAHS-A11, Analytical HPLC System LAHS-A10, and more.

Catalogs : Laboratory Equipment Catalog, Analytical Equipment Catalog, Medical Equipment Catalog, Test Equipment Catalog, Petroleum Testing Catalog, Pharmaceutical Machinery Catalog, Security Systems Catalog, Lab Products Catalog.

Need a Product

send details for best price quotation

Click
For

Labtron Equipment Ltd. Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom

Email: info@labtron.com

WhatsApp: +44 20 8004 3587

Phone: +44 2080 043608

Newsletter

Subscribe now to receive monthly news & personalised offer!

Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom

Email: info@labtron.com Whatsapp: +44 20 8004 3587 Phone: +44 2080 043608

Copyright © 2025 Labtron Equipment Ltd. All rights reserved.