Automatic Potentiometric Titrator LPTT-A18 features a measurement resolution of 0.1 mV, providing highly accurate detection of electrical potential changes during titrations. It includes multiple titration modes for precise control over titration. The titrator features chemically resistant tubing, providing durability against corrosive and sensitive reagents. Our automatic potentiometric titration system delivers a real-time display of the titration curve, ensuring precise monitoring and analysis throughout the process.
| Measuring Range | 2000.0 mV to +2000.0 mV -20.000 pH to +20.000 pH |
| Measurement Resolution | 0.1 mV |
| Measurement Precision | 0.1 mV ± 0.03% |
| Measurement Modes | MEAS: Measuring pH / mV/T CAL: pH correction (multi-point) SET Setting end-point mode MET Equivalent titrating DET Dynamic titrating |
| pH Resolution | 0.001 pH |
| pH Precision | 0.003 pH |
| Titrating Volume Accuracy | 0.001ml |
| Titrating Methods | Acid-base titration, redox titration, precipitation titration, complexation titration, non-water titration |
| Titrating Cup Capacity | 250 ml |
| Temperature Measurement Range | -5 to 120° C |
| Temperature Resolution | 0.01° C |
| Temperature Measurement Accuracy | ±0.1° C |
| Communication Mode | 1 USB, 232 serial port, ethernet |
| Fluid Adding Unit | Built-in liquid-adding unit |
| Liquid-Adding Module Unit | contains 1 sets of built-in titration liquid-adding channel. |
| Number of liquid-adding channels | 8 channels |
| Burette Volume | 1ml, 5ml, 10ml, 25ml, 50ml (10ml standard) |
| Burette Resolution | 1/48000 |
| Time of Replenishing Liquid | 10 sec |
| Control Unit | PC including software |
| Automatic Sampler | Compatible with GT robot injection system for mass sample analysis. |
| Self -Checking Function | Power-on to automatically check the instrument status. |
| Calibration Function | Yes |
| Storage Space | >300G (more than 10000 PCS) |
| Preset Method | >300G (more than 10000 PCS) |
| Pipelines | PTFE titration pipelines |
| PTFE Valve | Roof type spiral stirrer or magnetic stirring. |
| Microscale Detection Electrodes | Imported |
| Micro-Injection Pump | Imported |
Integrated Rotary Valve
Temperature Monitoring Mode
Automated Endpoint Detection
High-Capacity Storage
Overhead Stirring Coil
Automatic Potentiometric Titrator LPTT-A18 measures changes in the electrical potential to detect the endpoint. It is widely used in pharmaceutical, environmental testing, petroleum, and water treatment for accurately determining the concentration of active ingredients.
Download catalog for
Automatic Potentiometric Titrator LPTT-A18 Catalog
Available Range :
Automatic Potentiometric Titrator LPTT-A16 : -2000.0 to +2000.0 mV,Labtron Automatic Potentiometric Titrator are designed for precise potential measurement in the range of -1800.0 mV to +2000.0 mV. They ensure rapid filling and accurate burette volume control for efficient operation. Their linear motor transmission offers precise and controlled titrant delivery. They facilitate the storage of numerous titration results and allow easy customizations. Our table-top Automatic Potentiometric Titrator provide high-precision and cost-effective analysis in less time.








Labtron’s automatic potentiometric titrator LPTT-A18 operates by measuring the electrical potential (mV) between a reference and an indicator electrode as titrant is added to a solution. This method eliminates visual endpoint detection and instead relies on detecting inflection points in voltage. The automatic potentiometric titrator conducts real-time analysis and automatically determines the titration endpoint with high precision. It uses advanced algorithms to interpret titration curves and deliver accurate results. Ideal for redox, acid-base, precipitation, and complexometric titrations, this instrument enhances repeatability and consistency. It ensures operator-independent performance, minimizing human error. The system is widely used in pharmaceuticals, chemicals, food, and environmental testing labs.
Labtron’s automatic potentiometric titrator LPTT-A18 improves accuracy by automating titrant addition, electrode potential measurement, and endpoint detection. Unlike manual titrations, it eliminates parallax errors and subjective color change judgments. This automation ensures that each titration is carried out with identical parameters, improving repeatability. Sophisticated microprocessors in the automatic potentiometric titrator analyze titration curves and calculate exact concentrations. Temperature compensation and calibration protocols further ensure analytical precision. The closed system prevents contamination and minimizes sample exposure. Built-in data logging also allows traceability and validation of results. This makes it suitable for critical applications requiring stringent quality control.
Labtron’s automatic potentiometric titrator LPIT-A18 supports a wide variety of titrations, including acid-base, redox, precipitation, and complexometric titrations. It can also be used for non-aqueous titrations, especially in pharmaceutical and petrochemical analysis. For redox titrations, it monitors potential shifts due to electron transfer reactions. In acid-base titrations, it detects sharp voltage changes around the equivalence point. The automatic potentiometric titrator is also compatible with ion-selective electrodes for specific ions. Its flexible programming allows multi-step titrations and derivative curve analysis. Each titration type benefits from the instrument’s high sensitivity and automated control. This versatility makes it an essential lab asset.
Electrodes are central to the functioning of an automatic potentiometric titrator as they measure potential changes during titration. A reference electrode provides a constant potential, while an indicator electrode responds to analyte concentration. Proper maintenance includes regular calibration, cleaning with appropriate solutions, and avoiding mechanical damage. Storage in suitable electrolyte or buffer prevents drying out and degradation. The automatic potentiometric titrator often features electrode status monitoring to ensure optimal performance. Electrode lifespan depends on usage and care but can be extended with routine maintenance. Replacing worn electrodes ensures accurate and consistent results. Electrodes tailored to specific titration types enhance measurement accuracy.
The software in an automatic potentiometric titrator offers an intuitive interface for programming, monitoring, and analyzing titration procedures. It allows users to select titration methods, set parameters like titrant volume increments, and define endpoints. Real-time graphs of titration curves aid in understanding reaction progress. The software can perform statistical analysis and automatically generate reports. Built-in libraries help store standard methods and user-defined protocols for consistent operation. Advanced models may offer touchscreen controls, multi-language support, and connectivity with LIMS. Data export in multiple formats enhances documentation. The software ultimately enhances user productivity and analytical reliability.
Post your question?
Our experts are here to help!
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608