Request A Quote
Horizontal Laboratory Autoclave LHA-K15

Horizontal Laboratory Autoclave LHA-K15

Horizontal Laboratory Autoclave LHA-K15 features a robust design pressure of -0.1/0.25 MPa and a maximum working pressure of 0.23 MPa. It operates efficiently within a working temperature range of 105°C to 136°C and maintains exceptional temperature accuracy. It is equipped with a PLC control system, a color touch screen, and a built-in printer, ensuring user-friendly operation and accurate data recording. Its double-door structure enhances contamination control, while the included water tank and sterilization tray simplify workflow.

Horizontal Laboratory Autoclave LHA-K15
Horizontal Laboratory Autoclave LHA-K15
Horizontal Laboratory Autoclave LHA-K15
Horizontal Laboratory Autoclave LHA-K15
Request Quote
Capacity 200 L
Design Pressure -0.1 / 0.25 MPa
Max Working Pressure 0.23 MPa
Design Temperature 139°C
Working Temperature Range 105°C to 136°C
Temperature Accuracy ±0.1°C
Sterilization Time Range 0 to 99 min
Drying Time Range 0 to 99 min
Vacuum Rate -0.081 MPa
Door Type Double doors
Control System PLC + HMI + Built-in Printer
Display 7-inch LCD color touch screen
Vacuum System Pre-vacuum and post-vacuum
Steam Generator Built-in steam generator
Loading Type Sterilization tray
Water Tank Included
Drying Function Vacuum drying
Voltage 380V, 50/60Hz
Power Consumption 12 kW
Chamber Size 200L / Ø515 × 1000 mm
Outer Dimensions 1400 × 760 × 1750 mm
Packing Dimensions 1560 × 910 × 1880 mm
Net Weight 470 kg
Gross Weight 560 kg
  • Large Capacity Chamber
  • Wide Temperature Range
  • High Temperature Accuracy
  • Strong Pressure Handling
  • Advanced Vacuum System
  • User-Friendly Interface
  • Efficient Drying Function
  • Built-in Steam Generator
  • Double-Door Safety Design
  • Sterilization and Drying Time Adjustment
  • Integrated Water Tank
  • High Power Performance
  • Durable and Compact Design
  • Easy Loading System

Our Horizontal Laboratory Autoclave LHA-K15 is ideally suited for a wide range of sterilization needs in laboratories, hospitals, pharmaceutical industries, research institutions, and microbiology labs.

Models of Horizontal Autoclave

Labtron Horizontal Autoclave features side-loading capability with capacities ranging from 50 L to 2000 L. Their integrated steam generators provide a steady supply of sterilization vapors at temperatures from 105℃ to 150℃. Equipped with gauges for precise pressure monitoring and a display for checking pressure. The doors are safely interlocked, preventing accidental opening during operation. These Horizontal Autoclave efficiently sterilize medical and laboratory equipment by eliminating harmful microbes.

Horizontal Laboratory Autoclave LHA-B14

Horizontal Laboratory Autoclave LHA-B14

  • Capacity : 500 L
  • Sterilization Temperature : 134 °C
  • Sterilization Pressure : 0.22 MPa
  • Timer Range : 0 - 60 minutes
Horizontal Laboratory Autoclave LHA-F13

Horizontal Laboratory Autoclave LHA-F13

  • Capacity : 800 L
  • Sterilization Temperature : 139 °C
  • Sterilization Pressure : 0.25MPa
  • Steam Pressure and the Consumption : 0.4 - 0.6 MPa / 40 kg
Horizontal Laboratory Autoclave LHA-D23

Horizontal Laboratory Autoclave LHA-D23

  • Chamber Capacity : 150 Liters
  • Door Type : Double Door
  • Control System : PLC + HMI + Built-in Printer
  • Display : 7.0-inch Color Touchscreen
Horizontal Laboratory Autoclave LHA-E10
Horizontal Laboratory Autoclave LHA-E10
  • Capacity : 150 L
  • Sterilization Temperature : 134 °C
  • Sterilization Pressure : 0.22 MPa
  • Timer Range : 0 - 99 minutes
Horizontal Laboratory Autoclave LHA-I15
Horizontal Laboratory Autoclave LHA-I15
  • Capacity : 120 L
  • Working Temperature Range : 105 ℃ to 139 ℃
  • Sterilization Temperature : 150 ℃
  • Sterilization Pressure : -0.1/0.3Mpa
Horizontal Laboratory Autoclave LHA-B19

Horizontal Laboratory Autoclave LHA-B19

  • Capacity : 300L
  • Working Pressure : 0.22 MPa
  • Working Temperature : 134 °C
  • Temperature Range : 105 to 134 °C
Horizontal Laboratory Autoclave LHA-B10

Horizontal Laboratory Autoclave LHA-B10

  • Capacity : 150 L
  • Sterilization Temperature : 134 °C
  • Sterilization Pressure : 0.22 MPa
  • Timer Range : 0 - 60 minutes
Horizontal Laboratory Autoclave LHA-D12

Horizontal Laboratory Autoclave LHA-D12

  • Capacity : 600 L
  • Sterilization Temperature : 139 °C
  • Sterilization Pressure : 0.22MPa
  • Steam Pressure and the Consumption : 0.4 - 0.6 MPa / 30 kg
Horizontal Laboratory Autoclave LHA-K15 FAQ's
  • The Horizontal Laboratory Autoclave LHA-K15 operates within a precise temperature range of 105°C to 136°C, which is critical for effectively sterilizing a variety of laboratory, pharmaceutical, and hospital-grade instruments. This wide temperature range, paired with its ±0.1°C temperature accuracy, ensures that even heat-sensitive materials are sterilized without risk of under or over-processing. The ability to maintain this level of precision boosts sterilization reliability across different load types and volumes.

  • The LHA-K15 autoclave features a double-door safety design, which serves a vital role in controlled environments like cleanrooms or BSL labs. This design separates the "sterile" and "non-sterile" zones by allowing materials to enter through one side and exit through another after sterilization. Such a setup significantly reduces the risk of cross-contamination, especially in high-stakes environments like pharmaceutical production and hospital operating rooms.

  • The Horizontal Laboratory Autoclave LHA-K15 is equipped with a PLC control system integrated with an HMI interface and a 7-inch LCD color touchscreen. This intuitive control panel allows users to set, monitor, and adjust parameters like sterilization time, drying duration, and temperature with ease. Additionally, the built-in printer ensures that cycle data and sterilization logs are accurately recorded and documented for compliance and traceability.

  • Drying is a crucial post-sterilization step, and the LHA-K15 autoclave addresses this with its vacuum drying function, which removes residual moisture from instruments after steam sterilization. The advanced pre-vacuum and post-vacuum system enhances drying efficiency by evacuating air before sterilization and after the cycle ends, thus ensuring that instruments are sterile, dry, and ready for immediate use or packaging.

  • Yes, the Horizontal Laboratory Autoclave LHA-K15 is designed to meet high-throughput sterilization demands. It features a large 200-liter chamber with dimensions of Ø515 × 1000 mm, making it ideal for sterilizing bulky instruments, glassware, or multiple loads simultaneously. This makes the unit a reliable choice for hospitals, pharmaceutical industries, research labs, and microbiology facilities that require consistent and large-scale sterilization.

Post your question?

Our experts are here to help!

Need a Product

send details for best price quotation

Click
For

Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom

Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608

Copyright © 2026 Labtron Equipment Ltd. All rights reserved.