Elemental (CHN) Analyzer LCHN-A10 can handle sample masses up to 1000 mg. Offers flexibility in sample size depending on the matrix being analyzed. It has a stackable autoloader that can accommodate 35 samples in each layer. Utilizes a 5.0 ml blended gas analysis method to reduce helium and copper consumption. Incorporates IR CO2 and IR H2O detectors to ensure complete combustion and accurate measurement. This analyzer processes samples quickly, minimizing analysis time.
| Gas Required | He ≥99.995% O₂ ≥99.5–99.995% inert power gas |
| Measurement Range | Carbon: 0.005–100% Hydrogen: 0.01–60% Nitrogen: 0.008–60% |
| Consumption | Helium 200 ml/ min |
| Sample Mass | 80–100 mg recommended |
| Analysis Time | 4 to 6 mins, depending on sample combustion condition |
| Autosampler | 39 sample tray standard (optional 78 sample tray) |
| Accuracy | Carbon (C ): 0.45% Hydrogen (H ): 0.10% Nitrogen (N ): 0.05% |
| Furnace Type | Resistance furnace (main furnace and furnace rear) |
| Maximum temperature | 1050°C |
| Power Consumption | 4.5 KW |
| Power Supply | AC 220 ± 10 %V , 50/60 Hz |
| Dimension | 690 × 750 × 720 mm |
| Weight | 105 Kg |
Elemental (CHN) Analyzer is widely used in pharmaceuticals, chemicals, oil-related products, catalysts, and food industries for analysis of fuels, biomass, food, oil, coal, waste, plastic, solvents, fertilizers, etc.
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608