Specifications :
Ion Chromatographic pump |
Type | High pressure and low pulse double plunger tandem advection pump | Maximum pressure | 35 MPa (PEEK) | Display pressure accuracy | ≤ 0.1 MPa | Flow stability | (0.2 to 0.5) ml/ min ≤ 3 % , (0.5 to 1.0) ml/ min ≤ 2 %, > 1.0 ml/ min ≤ 2 % | Flow range | 0.001 to 9.999 ml/ min | Pressure pulse | ≤ 0.5 % |
|
Numerical control and electromagnetic injection |
Maximum pressure | 35 MPa | Contact material of the rotor | PEEK | Control mode | By Stepper motor | Power supply | DC 24 V |
|
Column heater |
Operating temperature range | Room temperature 20 to 60°C ( 68 to 140°F) | Controlling temperature accuracy | ± 0.01°C | Allowable deviation of column heater temperature | ± 2°C | Temperature stability | ≤ 1°C/ h |
|
Flow system |
Plastic flow path | Made up of PEEK material | Six way valve | PEEK material pressure 5000 psi, independent automatic collection and flow function |
|
Thermal buffer system for Eluent |
Temperature range | 25 to 40°C ( 77 to 104°F) |
|
Degassing device |
Type | Built-in low pressure | Vacuum degree | (-) 70 kPa | Maximum flow rate | 10 ml/ min | Internal volume | 30 µl | Degassing efficiency | 10 ml/ min 90 % |
|
Digital temperature and control detection system |
Type | Temperature-control and Bipolar conductivity detector | Detection mode | Bipolar conductivity detection | Cell volume | ≤ 0.8 µl | Detection range | 0 to 50000 µS/ cm | Output voltage | -6000 to +6000 mV (adjustable) | Operating temperature | Room temperature 5 to 60°C ( 41 to 140°F) | Controlling temperature accuracy | ± 0.01°C | Maximum pressure | 10 MPa | Linear range | ≥ 103 | Instrument linearity | ≥ 0.999 | Quantitative repeatability | ≤ 1.0 % | Qualitative repeatability | ≤ 1.0 % |
|
Eluent generator |
Eluent type | KOH/ NaOH | Eluent concentration range | 0.1 to 100 mM | Concentration increment | 0.1 mM | Flow rate change | 0.1 to 3.0 ml/ min | Maximum pressure | 20 MPa | Minimum pressure | 5 MPa |
|
Power |
150 W |
Power supply |
DC 220 V, 60 Hz |
External dimension |
500 × 500 × 760 mm |
Net weight |
48 kg |
Gross weight |
73 kg |
HS code |
9027201200 |
Applications :
Used to detect small molecules (in the form of anion, cation) in water testing & processing plants, food, chemical and dye industry, mining, metallurgy, and other fields.
Optional Accessories :
Accessories no. |
Name |
1 |
Anion column |
2 |
Guard column |
3 |
Anion Suppressor |
4 |
UV detector |
5 |
Ampere detector |
6 |
Auto sampler (108) samples |
Auto Sampler :
Display |
7 inch touch control, color LCD display |
Injection mode |
X-Y-Z three axis operational mode |
Sampling mode |
Full quantitative loop sampling |
Number of injection units |
2 ml × 120 units |
Repeat injection |
1 to 5 times (optional) |
Injection volume |
25 to 500 µl |
Sample syringe capacity |
5 ml maximum |
Pipe material |
Plastic |
Circulation time |
≤ 6 minutes |
Injection accuracy |
≤ 0.5 % |
Residual sample |
≤ 0.05 % |
Control system |
ARM STM32 series of 32 bit embedded |
Embedded |
LQFP100 |
Communication mode |
RS232 |
Dimension (L×W×H) |
400 × 540 × 260 mm |