IR Carbon and Sulphur Analyzer LCSA-A13 stands out with its exceptional detection sensitivity of 0.1 ppm. It is achieved through a combination of a precious metal light source, pyroelectric solid light cone sensor, and narrowband filters. Great stability is ensured by stepper motor control, modular circuit design and 32-bit sampling chip. Boasts military-grade power modules, a special aviation motor, and a ceramic high-power tube for durability. This comprehensive solution provides precise results, unwavering reliability, and user-friendly operation.
| Measurement Range (Carbon) | 0.0001% to 15.0000% (expandable to 99.9999%) |
| Measurement Range (Sulfur) | 0.0001% to 10.0000% (expandable to 99.9999%) |
| Sensitivity | 0.1 ppm |
| Analysis Time | Adjustable within 25 to 60 s (Typically 35 s) |
| Working Environment | Indoor Temperature: 10 to 30 ℃ Relative Humidity: < 75% |
| Working Gas | Oxygen, Purity > 99.5% |
| Analysis Error | Meets ISO9556 (Carbon), ISO4935 (Sulfur), JJG395-97 standards |
| Oscillating Frequency | 20 MHz |
| Combustion Power | 2.5 KVA to 7.5 KVA (auto-adjust) |
| Operating Environment | Temperature: 10 °C to 30 °C, Humidity: ≤ 75% |
| Analysis Room Area | Requires more than 3 × 3 m |
| Power supply | AC 220 V ± 5%, 50 Hz ± 2 |
| Dimensions | 1420 × 770 × 1100 mm |
| Weight | 210 Kg |
| Accessories No. | Name | Specification |
| 1 | Ceramic Crucible | 1000 Pcs/box |
| 2 | Tungsten Accelerator | 1000 g/bottle |
| 3 | Pure Iron Accelerator | 100 g/bottle |
| 4 | Desiccant | 50 g/bottle |
| 5 | Silicon Ring | - |
| 6 | Steel Wire Brush | - |
| 7 | Quartz Combustion Tube | - |
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608