Liquid Scintillation Counter LSCL-A10 can handle up to 260 samples. It operates in energy spectrum analysis mode to assess sample properties. 20ml standard bottles are used for samples. The system employs 3+3 coincidence and anti-coincidence detection and the TDCR quench correction for accuracy. It supports both relative and absolute measurements that can be scaled without a standard source. Our liquid scintillation counter is programmable for unattended measurements, allowing multiple tasks to run simultaneously.
| Number of Samples | 260 |
| Analysis Mode | Energy spectrum |
| Sample Feeding Method | Automatic conveyor belts |
| Sample Containers | 20ml standard bottle |
| Multi-channel Analyser | 2 × 2048 channels |
| Temperature Control Devices | System built-in |
| Energy Range | α: 3 to 10 MeV; β: 0 to 5 MeV |
| Recovery Efficiency | 14C: more than equal to 95% |
| Background | less than 2.0 cpm (20 ml containing 40% water,3H efficiency>25%), 3H (0 to 18.6 keV<20 cpm;14C (0 to 156keV) less than 25cpm. |
| Lower Detection Limit | 1.5 Bq/L (12 ml scintillation solution + 8 ml tritium water sample measured for 1000 min) |
| Memory 14C | Depends on the equipped computer |
| Standard Storage | 2GB |
| Hard Drive | 320GB |
| Built-in Storage | Not available (Requires connection to a computer) |
| Consumables | 10L Ultima GOLD LLT Scintillation Fluid Unquenched standard source (One bottle each of 3H, 14C and Background) 3H Series Quench Standard Source 14C Series Quench Standard Source |
| 24h Instability | Count change less than 0.2%/24h |
| Energy Resolution | 0.01 keV/ch (³H) |
| Operating Temperature | 5°C to 35°C |
| Operating Humidity | 30% to 80% (25°C, no frosting) |
| Communication Methods | USB, RJ45 |
| Power Supply | AC 220V ±10%, 50Hz ±10% |
| Power | Mainframe 200W, temperature control system 500W |
| Dimension | 1245 × 900 × 1145 mm |
| Packaging Dimension | 1300 × 1020 × 1450 mm |
| Net Weight | 380 Kg |
| Gross Weight | 520 Kg |
Dual multi-channel analysis technology
Supports double marker separation
Provides automatic sample exchange
Advanced data processing
Logarithmic and linear spectrograms
Liquid Scintillation Counter LSCL-A10 accurately measures low levels of ³H and ¹⁴C and monitors radioactive effluents from nuclear power plants.
Accessories Name | Specification | Quantity (Pcs/set/ bottle) |
10-position sample holder | Auto sampler | 26 pcs |
Control & analysis software | Spectral acquisition and analysis | 1 set |
All-in-one PC | 23” | 1 set |
Power line | 1.5m | 1 pcs |
Polyethylene bottle | 20ml | 20 pcs |
Samplers and pipettes | 5ml | 1 pc |
Antistatic fluid | 500ml | 1 bottle |
Consumable Names | Quantity |
20 ml Polyethylene Bottle | 1000 pcs |
20 ml Low Potassium Glass Bottle | 500 pcs |
20 ml PTEE Coated Plastic Bottle | 100 pcs |
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom