Labtron Cleveland Open Cup Flash Point Tester LOFP-A11 is designed for accurate flash point testing of oils, petroleum products, and lubricants. It offers a temperature range up to 400°C with 0.1°C resolution for accurate measurements. The tester uses the Cleveland open cup method and features electronic ignition with an adjustable heating rate. The cleveland flash point tester is suitable for petroleum testing, lubricant analysis, fuel testing, and laboratory quality control. The cleveland open cup apparatus provides consistent results for routine flash point testing in laboratory and industrial applications.
| Temperature Measurement Range | Room Temperature to 400℃ |
| Repeatability | Flash Point ≤150℃: ±2℃ Flash Point ≥150℃: ±3℃ |
| Resolution | 0.1℃ |
| Precision | 0.5% |
| Heating Speed | In line with GB/T3536 ASTM D92 Heating rate: 5 to 17℃ per min |
| Ignition Mode | Electronic ignition, flame size 3mm to 4mm |
| Ambient Temperature | 10 to 40℃ |
| Relative Humidity | ≤80% |
| Power Supply | AC 220V ±10%, 50Hz |
| Packaging Dimensions (L × W × H) | 350 × 530 × 400 mm |
| Net Weight | 15 kg |
| Gross Weight | 19 kg |
Our Cleveland Open Cup Flash Point Tester LOFP-A11 is mainly used to determine the flash point of various types of oils, including gasoline engine lubricating oil, diesel engine lubricating oil, compressor oil, refrigeration oil, turbine oil, gear oil, and mechanical oil.
| Accessories Name | Quantity |
| Automatic opening flash point meter | 1 set |
| Oil cup | 1 piece |
| AC power cord | 1 piece |
| Printing paper | 1 piece |
| Standard Test Crucible cup with Handle | 1 piece |
| Temperature probe | 1 piece |
| Electric Igniters | 1 piece |
| Static Relays in the device | 1 piece |
| Heating Element | 1 piece |
Certified reference materials with flash point minimum 200 °C, maximum 240 °C (250mL)
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608