Overview
Semi-Automated Electrolyte Analyzer LSAE-A13 can process up to 60 samples per hour, supporting both primary tubes and cuvettes. Liquid level detection and probe collision protection improve reliability during auto-sampling. With rapid testing capability, completing ISE tests in just 25 seconds, it ensures high throughput. Its electrodes provide long lifespan, anti-jamming, and anti-pollution properties. Our analyzer with a lightweight mainframe and samplers, allows for easy integration into various laboratory settings.
Specifications :
| Sample Type |
Plasma, Serum, Whole blood, Urine, CSF |
| Test Parameters |
K⁺, Na⁺, Cl ⁻, Ca²⁺, Li⁺, pH, TCO₂ |
| Sample Volume |
65 to 150µL |
| Test Range |
K⁺: 00.50 to 15.00 mmol/L
Na⁺: 20.0 to 200.0 mmol/L
Cl ⁻: 20.0 to 200.0 mmol/L
Ca²⁺: 0.10 to 5.00mmol/L
Li⁺: 0.00 to 3.00mmol/L
pH: 6.00 to 9.00
TCO₂: 6.0 to 50.0 mmol/L |
| Test Resolution |
K⁺: 0.01 mmol/L
Na⁺: 0.1 mmol/L
Cl ⁻: 0.1 mmol/L
Ca²⁺: 0.01mmol/L
Li⁺: 0.01mmol/L
pH: 0.01
TCO₂: 0.1 mmol/L |
| Test Precision |
K⁺: ≤ 1.0%
Na⁺: ≤ 1.0%
Cl ⁻: ≤ 1.0%
Ca²⁺: ≤ 1.5%
Li⁺: ≤ 1.5%
pH: ≤ 1.0%
TCO₂: ≤ 3.0% |
| Sample Load |
40 Samples/Hour (with TCO₂)
60 Samples/Hour (without TCO₂) |
| Sample Loader |
25 Positions
40 Positions |
| Sample Tube |
Primary tube or cuvette |
| Ambient Temperature |
5℃ to 40℃ |
| Relative Humidity |
≤ 85% |
| Power Supply |
100 to 240V, 50/60 Hz |
| Small Sampler Dimension |
347 × 277 × 451 mm |
| Big Sampler Dimension |
570 × 425 × 380 mm |
| Mainframe Dimension |
337 × 192 × 402 mm |
| Small Sampler Weight |
6.5 kg |
| Big Sampler Weight |
10 kg |
| Mainframe Weight |
17 kg |
Features :
- Self-diagnostic and self-adjustment
- Alarm for low reagent volume
- Real-time testing status
- STAT capability for single/multiple sample tests
- Liquid level detection
- Probe collision protection
- Built-in bar code scanner
- All-in-one maintenance mode
- Visible QC graph for storage, transfer, and printing
Applications :
Our Semi-Automated Electrolyte Analyzer is used in clinical laboratories to measure and analyze key electrolytes such as potassium, sodium, and chloride in samples like plasma, serum, whole blood, urine, and cerebrospinal fluid for diagnostic purposes.
Standard Accessories :
Electrode set
Internal thermal printer
External sample bar code scanner
Auxiliary pump tube × 2
Pump tube (2.4 × 5.6 BPT) × 3
Reagent pack connector
Allen key
Fuse × 2
Printer paper (30 × 57) × 3
Electrode O ring (Φ2.6 × 1.9) × 5
Syringe (50ml)
Taike tube (G2.5 × 0.8) 200mm
Long needle for blockage elimination (Φ0.6 × 160)
Manual-probe tie in × 2
Bottle cap (with a small hole) for TCO2 Reaction Solution
Tubes for Auto sampler (12.7mm × 100mm)
Tubes for Auto sampler (12.7 × 75mm)
Cuvettes for Auto sampler
Auto Probe tie in for Auto sampler × 2
Data cable for Auto sampler
Refill Solution for ISE Electrodes (0.8ml) × 5
Refill Solution for Reference Electrode (20ml)
Calibration standard Solution
De-proteinizer
Electrolytes Quality Control (High/ Medium/ Low)
Conditioning Solution (0.8ml) × 5
TCO₂ Reagent (50ml) × 3
TCO₂ Quality Control (H:1ml×5; L:1ml×5)
Optional Accessories :
Small auto sampler (25 sample positions)
Small auto sampler (25 sample positions) with built-in bar code scanner
Big auto sampler (40 sample positions) with built-in bar code scanner K electrode
Na electrode
Cl electrode
Ca electrode
pH electrode
Li electrode
Ref electrode
Refill solution for Ref electrode
Refill solution for ISE electrodes (K/Na/Cl/Ca/pH)
Refill solution for Li electrode
Cleaning solution
Conditioning solution
Calibration Standard Solution for ISE
Calibration Standard Solution for ISE + Li
Reagent pack connector
TCO₂ Reagent- Reaction solution (50ml)
TCO₂ quality control (High/Low)
De-proteinizer (including diluter) for electrodes
Electrolytes quality control (High/Medium/Low)
Selective coefficient calibrator for Li electrode
Pump tube
Printer paper
Auto-Probe tie in
O’ ring for electrode
Cuvettes for auto sampler × 1000