Temperature and Humidity Test Chamber LTHC-I12 offers an atmospheric pressure range of 80 kPa to 106 kPa for optimal performance. The controller includes a built-in data logger, enabling long-term data recording and export to CSV format. It features energy-saving mechanisms such as automatic compressor cutoff during stable operation. Its structure includes an SUS304 stainless steel inner chamber and a cold-rolled steel exterior with electrostatic coating. Our chamber supports accelerated aging, moisture sensitivity, and thermal cycling tests.
| Capacity | 225 L |
| Temperature Range | -40℃ to +150℃ or -70℃ to +150℃ |
| Humidity Range | 20% to 98% RH |
| Temperature Fluctuation | ±0.5℃ |
| Humidity Fluctuation | ±1% RH |
| Temperature Deviation | Less than or equal to ±1℃ |
| Humidity Deviation | ±3% RH |
| Temperature Uniformity | Less than or equal to 2.0℃ (-40℃ to +150℃ range) |
| Heating Rate | more than or equal to 3℃/minute |
| Cooling Rate | more than or equal to 1℃/minute |
| Heating & Cooling Overshoot | Less than or equal to 2℃ |
| Noise Level | Less than or equal to 70 dB (A sound level) |
| Ambient Temperature | 5℃ to 35℃ |
| Relative Humidity | Less than or equal to 85% RH |
| Atmospheric Pressure | 80 kPa to 106 kPa |
| Frequency | 50 ± 0.5 Hz (60 Hz supported) |
| External Chamber | Electrostatic spraying of cold rolled steel plate |
| Inner Chamber | SUS304 stainless steel plate |
| Insulating Layer | Hard polyamine foam + ultra-fine eco-friendly glass fiber cotton |
| Observation Window | 3-layer vacuum large glass (30 × 40 cm), with electric heating anti-frost/condensation function |
| Test Hole | 50 mm or 100 mm on the left side (Standard: 1 piece) |
| Sample Racks | 2 pieces |
| Pressure Balance Port | 1/2 inch drainage shared type |
| Light | 9W LED energy-saving lamp mounted at observation window |
| Test Standards Compliance | IEC60068-2-1:2007 (Low temperature test) IEC60068-2-2:2007 (High temperature test) IEC60068-2-78:2007 (Test Cab: Steady state damp heat test method) IEC60068-2-30:2005 (Test Db: Alternating damp heat test method) |
| Power Supply | AC 380V 5.5KW |
| Chamber internal size | 500 x 600 x 750 mm |
| Chamber external size | 750 x 1500 x 1850 mm |
| Maximum Load Capacity | 20 kg (chamber bearing) |
| Chamber weight | 300 kg |
Wide Temperature Range
Adjustable Humidity Control
Energy-Efficient Compressor
Multi-Layer Insulation Design
Stainless Steel Chamber
Our Temperature and Humidity Test Chamber supports accelerated aging, moisture sensitivity, and thermal cycling tests for semiconductors and PCBs, ensuring compliance and enhancing product durability.
USB Interface
RS-232
Crystal Head
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608