Diffraction Grating Spectrophotometer LDGS-A10 features the sphere size of Φ 48mm, enhancing light capture and accurate reflectance. It offers multiple detector options, ensuring for the selection of the ideal detector based on sample types. The instrument supports various color spaces for accurate color measurements across different materials. Our diffraction grating spectrophotometer has polygon colorimetric tolerance for adjusting temperature variations.
| Optical Geometry | 45/0(45°Ring Shaped Illumination, Vertical Viewing) |
| Sphere Size | Φ48 mm |
| Sensor | 256 Image Element Double Arrays CMOS Image Sensor |
| Wavelength Range | 400 to 700 nm |
| Wavelength Pitch | 10nm |
| Half Bandwidth | 10nm |
| Reflectance Range | 0 to 200% |
| Measurement Aperature | Φ 20 mm |
| Color Spaces | CIE LAB, XYZ, Yxy, LCh, CIE LUV, Hunter LAB, Bxy |
| Color Difference Formula | ΔE × ab ΔE × uv, ΔE × 94, ΔE × cmc (2:1), ΔE ×cmc (1:1), ΔE × 00, ΔE (Hunter) |
| Other Indices | WI (ASTM E313, CIE/ ISO, AATCC, Hunter) YI (ASTM D1925, ASTM 313) MI (Metamerism Index) |
| Observer Angle | 2°/ 10° |
| Illuminants | D 65, A, C, D50, D55, D75, F1, F2 (CWF), F3, F4, F5, F6, F7 (DLF), F8, F9, F10 (TPLF), F11(TL84) F12 (TL83/ U30) |
| Measurement Time | 1.5 second |
| Repeatability | Spectral Reflectance: MAV / SCI Standard Deviation: 0.1% (400 to 700 nm) Chromaticity Value: Standard Deviation ΔE × ab 0.04 |
| Inter – Instrument Agreement | ΔE × ab 0.2 |
| Data Storage | Standard: 1000 Sample: 30000 |
| Measurement Mode | 2 to 99 times |
| Battery Performance | 5000 Measurement within 8 hrs |
| Working Environment | Temperature: 0 to 40°C Humidity: 0 to 85% (No Condensation) Altitude: Less than 2000 m |
| Storage Environment | Temperature: 20 to 50°C Humidity: 0 to 85% (No Condensation) |
| Power Consumption | DC 24V, 3A |
| Dimensions (L× W× H) | 184 × 77 × 105 mm |
| Weight | 600g |
PC Software Integration
Reflective Film Evaluation
Standard White Calibration
Multiple Colorimetric Indices
Minimizing Stray Light
Diffraction Grating Spectrophotometer LDGS-A10 is an analytical instrument that uses a grating to split light into wavelengths for accurate analysis. It is widely used in the pharma and laboratories, ensuring precise color in tablets and packaging for quality control.
Accessories Name | Quantity |
Power Adaptor | 1 |
USB Cable | 1 |
Built-in li-ion Battery | 1 |
Protective Cover | 1 |
White and Black Calibration Cavity | 1 |
Management Software | 1 |
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom