Portable Dry-Block Temperature Calibrator LPDTC-A10 operates within a range of -25°C to 150°C for accurate calibration. It maintains temperature stability of ≤±0.05°C per 15 minutes, ensuring consistent performance. A 150mm insertion depth allows sufficient immersion for sensor calibration. The digital control system enables rapid heating and cooling adjustments for efficient operation. Its compact, lightweight design makes it ideal for on-site calibration in laboratories and industries.
| Temperature Range | -25°C to 150°C |
| Display Resolution | 0.01°C (optional 0.001°C) |
| Temperature Deviation Accuracy | ±0.1°C |
| Temperature Stability | equal to or less than ±0.05°C/15min |
| Horizontal Temperature Uniformity | equal to or less than ±0.1°C |
| Vertical Temperature Uniformity | equal to or less than 0.5°C |
| Insertion Depth | 150mm (customizable) |
| Soaking Block Aperture | Standard 5 holes (Ø4, Ø6, Ø6, Ø8, Ø10)mm (customizable) |
| Outer Diameter of Soaking Block | Ø30mm |
| Power | 500W |
| Power Supply | 220VAC; 50Hz |
| Overall Dimensions | 365×190×400 mm |
| Weight | 8.5kg |
High-brightness backlit display
Uniform heat distribution
Replaceable soaking blocks for flexibility
Overheat and safety protection system
Low power consumption
Multi-sensor compatibility support
Integrates with external calibration software
Top-mounted handle for easy carrying
Our Portable Dry-Block Temperature Calibrator LPDTC-A10 ensures precise calibration of thermocouples, resistance thermometers, and temperature transmitters. Ideal for industrial, research, and metrology applications, its portable design enables on-site calibration.
Soaking Block Handle
Soaking Block
Power Cable
Labtron Equipment Ltd., Quatro House, Lyon Way, Camberley, Surrey GU16 7ER United Kingdom
Email: info@labtron.com Whatsapp: +44 73 4441 2688 Phone: +44 2080 043608