Liquid Chromatography Mass Spectrometry LCMS-A11 features a flow rate range of 1 to 4000 µL/min, adjustable in 1.0 µL/min steps for precise control. It uses a triple quadrupole tandem mass spectrometer for accurate analysis. The system supports full loop, semi-loop, and microliter injection modes to suit various sample needs. It is equipped with E Spray biorthogonal electrospray ion source technology, offering high ionic yield and strong resistance to matrix effects. Our liquid chromatography mass spectrometry delivers high sensitivity in detecting sulfonamides in pork, pesticides in water, environmental pollutants, neonatal diseases, and vitamin D in serum.
| Flow Range | 1 to 4000 µL/min, increments of 1.0 µL/min |
| Accuracy of Flow Rate | ≤1% |
| Precision of Flow Rate | ≤0.07% RSD |
| Maximum Pressure | ≥18000 psi |
| Mobile Phase Selection | Solvent selection valve allows any two of A, B, C, D as mobile phase |
| Vacuum Degasser | Built-in; each pump degassed separately (A / B) |
| Delay Volume | <50 µL (includes 35 µL mixer) |
| Gradient Modes | Linear, concave, convex |
| Gradient Range | 0 to 100% |
| Gradient Precision | ≤0.15% RSD |
| Gradient Accuracy | ±0.5% (absolute), 5% to 95% |
| Pressure Fluctuation | <1% of system pressure or 5 bar (whichever greater) |
| Injection Modes | Full loop, semi-loop, microliter |
| Injection Repeatability | Full: <0.3% RSD; Semi: <0.5% RSD; Micro: <1.0% RSD |
| Sample Capacity | Max 384 positions; ≥96 positions for 2 mL vials |
| Injection Range | 1–500 µL (standard); 250, 1000, 2500 µL (optional) |
| Residue | <0.05% (standard); <0.01% (extra rinse); |
| Sample Cooling | 4℃±2℃ (min); Room temp to -3℃(max) |
| Temperature Control Range | Room temp +5℃ to 90℃ |
| Temperature Control Method | Mobile phase preheating + forced air circulation |
| Maximum Column Capacity | ≥5 columns, length ≥250 mm |
| Design | Orthogonal vertical spray, anti-pollution, low background noise |
| Ion Source Gases | 1 atomizing + 2 desolvation gases; flow rate set in software |
| Spray Voltage | Max 6 kV; settable in software |
| Aux Heating Temp | ≥700℃; settable in software |
| Exhaust Device | Prevents backflow, reduces memory effect, protects pump and environment |
| Ion Interface | Cone hole or capillary; capillary design requires 50 extra capillaries |
| Back-Blowing Design | Reduces neutral molecule introduction |
| Interface Temp | ≥110℃ to improve anti-pollution |
| Ion Transmission System | 4-stage differential vacuum with multi-quadrupole focusing |
| Mass Analyzer | Triple quadrupole |
| Quadrupole Material | High-precision pure Mo; no need for extra temp control |
| Quadrupole Surface | Gold-plated for anti-contamination |
| Quadrupole Structure | Pre-rods before/after first; front pre-rod in second quadrupole |
| Collision Cell | Linear or 90° curved; curved requires 3 years free cleaning commitment |
| Collision Gas | High-purity nitrogen |
| Sensitivity | ESI+: 1 pg reserpine, S/N≥300,000:1; ESI-: 1 pg chloramphenicol, S/N≥300,000:1 |
| Quantitative Reproducibility | Reserpine & Chloramphenicol: RSD ≤5% (6 injections) |
| Quadrupole Resolution | 0.4 to 3.0 amu adjustable |
| Mass Stability | ≤0.1 amu/24h |
| Scan Speed | ≥20,000 amu/s |
| Min Scan Step | 0.02 amu |
| Dynamic Range | 6 orders of magnitude |
| MRM Channels | ≥30,000 MRM groups per injection with no sensitivity loss |
| Min Dwell Time | 1 ms |
| MS/MS Switch Time | ≤1 s |
| Detector Type | Electron multiplier tube |
| Detection Method | Pulse counting for low concentration reproducibility |
| Vacuum System Configuration | Mechanical + turbomolecular pump, differential exhaust, auto power-off |
| Mechanical Pump Speed | 65 m³/h |
| Turbo Pump Speed | 50 / 400 / 500 L/s |
| Workstation Software Platform | Windows XP and above |
| Automation | Calibration, instrument monitoring |
| Power Supply | (220±20) V AC,10A,50 Hz |
| Dimension | 750 × 660 × 670 mm |
| Packaging Dimension | 940 × 830 × 600 mm |
| Net Weight | 48 Kg |
| Gross Weight | 68 Kg |
Step scan ion transmission
Low-noise pulse counting detection
Superior stability
High sensitivity
User-friendly design
Liquid Chromatography Mass Spectrometry LCMS-A11 detects sulfonamides in pork, pesticides in water, pollutants, neonatal diseases, and vitamin D levels in serum with high sensitivity.
Accessories Name | Quantity |
Ultra-High Pressure Gradient Pump | 1 pc |
Automatic Sample Injector | 1 pc |
Column Oven | 1 pc |
ESI Ion Source | 1 pc |
Vacuum Interface and Ion Transmission System | 1 pc |
Mass Analysis System | 1 pc |
Detector | 1 pc |
Vacuum System | 1 pc |
Workstation Software | 1 pc |
Accessories Name | Quantity |
Computer System | 1 pc |
AC Regulated Power Supply | 1 pc |
Laser Printer | 1 pc |
Nitrogen Generator | 1 pc |
Bottle of High-Purity Nitrogen (including gas pressure reducing valve) | 1 pc |