Automatic COC Flash Point & Fire Point Tester LOFP-B11 measures ignition of liquid vapor and its burning temperature from 79℃ to 370℃. It features fast heating and cooling cycles enhance testing efficiency for accurate results. Integrated with pluggable igniter for convenient and easy replacement. Built-in safety mechanism protects operators during hazardous condition. Our Tester is ideal for research, quality control, and production monitoring.
| Measuring Range | 79 to 370℃ (175 to 698℉) |
| Resolution | 0.1℃ |
| Repeatability | ≤ 6℃ |
| Reproducibility | ≤ 8℃ |
| Gas Supply | LP gas, butane gas, pipeline gas, or natural gas (pressure < 10kPa) |
| Cooling | Air cooling |
| Ignition Source | Gas ignition with automatic lighting |
| Flash Point Detection | Ionization Ring |
| Temperature Sensor | PT100 probe |
| User Interface | 7 |
| Supported Languages | Chinese, English, Russian |
| Barometric Correction | Flash point corrected to barometric pressure |
| Data Storage | No less than 5000 results |
| Standards | ASTM D92, ISO 2592, IP 36, GB/T 3536 |
| I/O Ports | RS232, RJ-45 |
| Safety Features | Automatic shutdown, alarm, and cooling under: Flash Point Mode: EFP + 20℃ Flash & Fire Mode: EFP + 50℃ Overheating: 400℃ |
| Data Export | LIMS or Printer |
| Machine Power | 900W |
| Heater Power | 800W |
| Operating Conditions | 5℃ to 35℃ |
| Relative Humidity | 20 to 90% , 35℃ (non-condensing), no airflow |
| Power Supply | AC 220V ± 10%, 50/60Hz |
| Dimensions (W x D x H) | 530mm x 233mm x 427mm |
| Weight | 19 Kg |